5.3.6If Fuse blows, connect the unit through a variable AC Source with a current meter and monitor the current at no load. If the current is very high (more than 2 ampere or so ) load, check for Bridge Rectifier short or shorted Diode, or Secondary Winding short, etc
5.3.7If output voltage or current DPM's are not reading, check for loose connections especially in crimping, soldering, of Connectors.
WARNING
THE FOLLOWING SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY.TO AVOID PERSONAL INJURY, DO NOT PERFORM ANY SERVICING OTHER THAN THAT CONTAINED IN
-OPERATING INSTRUCTIONS UNLESS YOU ARE QUALIFIED TO DO SO.
REFER TO OPERATORS SAFETY SUMMARY AND SERVICE | SUMMA |