Agilent Technologies 668xA Power Supply Hardware Error Messages, Calibration Error Messages

Page 59

5

Error Messages

Power Supply Hardware Error Messages

Front panel error messages resulting from selftest errors or runtime failures are described in “Chapter 3 - Turn-On Checkout” of the power supply Operating Guide.

Calibration Error Messages

Front panel error messages resulting from calibration errors are described in Appendix A of the power supply Operating Guide.

System Error Messages

System error messages are obtained remotely with the SYST:ERR? query or by pressing the front panel key. The error number is the value placed in the error queue. SYST:ERR? returns the error number into a variable and combines the number and the error message into a string. Table 5-1 lists the system errors that are associated with SCPI syntax errors and with interface problems. Information inside the brackets is not part of the standard error message, but is included for clarification. When system errors occur, the Standard Event Status register (see "Chapter 4 - Status Reporting") records them as follows:

Standard Event Status Register Error Bits

 

Bit Set

Error Code

Error Type

Bit Set

Error Code

Error Type

5

-100 thru -199

Command

3

-300 thru -399

Device-dependent

4

-200 thru -299

Execution

2

-400 thru -499

 

Query

 

 

Table 5-1. Summary of System Error Messages

 

 

 

Error

Error String [Description/Explanation/Examples]

 

 

 

Number

 

 

 

 

 

 

 

-100

Command error [generic command error]

 

 

 

 

 

-101

Invalid character

 

 

 

 

 

 

-102

Syntax error [unrecognized command or data type]

 

 

 

 

 

-103

Invalid separator [illegal character encountered in place of separator]

 

 

 

-104

Data type error [e.g., “numeric or string expected, got block date”]

 

 

 

 

-105

GET not allowed [ <GET> inside a program message]

 

 

 

 

-108

Parameter not allowed [too many parameters]

 

 

 

 

 

-109

Missing parameter [too few parameters]

 

 

 

 

 

-112

Program mnemonic too long [maximum 12 characters]

 

 

 

 

-113

Undefined header [syntactical correct but not defined for this device]

 

 

 

-121

Invalid character in number [e.g. alpha in decimal data, etc.]

 

 

 

 

-123

Exponent too large [ numeric overflow; exponent magnitude >32000]

 

 

 

-124

Too many digits [number too long; more than 255 digits received]

 

 

 

 

-128

Numeric data not allowed [numeric data not accepted where positioned]

 

 

 

-131

Invalid suffix [unrecognized suffix, or suffix not appropriate]

 

 

 

 

-138

Suffix not allowed [numeric element does not allow suffixes]

 

 

 

Error Messages 59

Image 59
Contents Agilent Part No Microfiche Part No July Programming Guide Gpib DC Power SuppliesPrinting History Safety GuidelinesContents Display Subsystem Description of Subsystem CommandsSystem Commands Current SubsystemQuestionable Status Group Command Summary Programming ParametersPower Supply Status Structure Operation Status Group Voltage SubsystemPage User’s Guide General InformationAbout this Guide Documentation SummaryAccessing Online Help Prerequisites for Using this GuideVXIplug&play Power Product Instrument Drivers Downloading and Installing the DriverConventions Gpib Capabilities Of The Power SupplyRemote Programming Introduction To ScpiVoltlev 4.5PROT 4.8CURR?NL Types of Scpi CommandsScpi Messages Structure of a Scpi MessageVoltlev 4.5 Prot 4.8 CURR? Parts of a Scpi MessageVolt LEV Prot Curr Root Specifier Traversing the Command TreeQuery Indicator Message Unit SeparatorOUTPUTPROTECTIONCLEARSTATUSOPERATIONCONDITION? Effect of Optional HeadersMoving Among Subsystems Outputprotectionclear STATUSOPERATIONCONDITION?Scpi Data Formats Including Common CommandsScpi Queries Value CouplingExamples Controlling the Output Programming Status Saving and Recalling StatesWriting to the Display Digdata System ConsiderationsProgramming the Digital I/O Port Gpib AddressAssigning the Gpib Address In Programs Action Display ShowsSample Program Code Error HandlingDOS Drivers Agilent Basic ControllersIout = OUTPUTS2 Programming Some Power Supply FunctionsCall Ioenter PS,OEVENT SPOL%=O While C O Introduction Language DictionaryCommon Commands Syntax Diagram Description Of Common CommandsESE CLSOPC ESR?IDN? PSC 0 *PSC OPC?OPT? PSCPSC *RST *SAV Command Syntax RCL NRf Parameters ExampleQuery Syntax None Related Commands RCLSAV RSTSAV SRE Bit Configuration of Status Byte RegisterSRE STB?WAI TRGTST? Abor Description of Subsystem CommandsCalibration Commands AborCurrprotstat Current SubsystemCurr Currtrig Digdata 7 Digitaldatavalue Display SubsystemDigdata DispDispmode Norm Displaymode Normal Displaywindowmode Text Disptext DefaultmodeDispmode DisptextMEASCURR? MEASVOLT? Initiate SubsystemMeasure Subsystem Init InitcontOutpprotcle Outpprotdel Output SubsystemOutp Outprelpol Norm OutprelOutprelpol Outprel 1 Outprel OFFSTATOPER? Status SubsystemStatpres Status Operation RegistersStatoperenab 1312 Statoperenab STATUSOPERATIONENABLE?Statoperenab Statoper NTR Statoper PTRStatquesenab Status Questionable RegistersSTATQUES? STATQUESCOND?SYSTERR? System CommandsSYSTERR? SYSTEMERROR? Statques NTR Statques PTRTrig Trigger SubsystemSystlang SYSTVERS?Trigsour BUS Triggersource BUS Voltage SubsystemTrigsour Volt VolttrigVoltprot Command SummaryCommand Summary Command Parameters Subsystem CommandsCommon Commands Parameters Command ParametersVoltprot MAX Programming ParametersCurrlevtrig MAX Voltlevtrig MAXPage Operation Status Group Power Supply Status StructureRegister Commands Status ReportingOPC CALWTG UNRSTATQUESNTR? Status Questionable Commands Register Query Cleared ByQuestionable Status Group Standard Event Status GroupOutput Queue Service Request Enable RegisterInitial Conditions At Power On Status Byte RegisterStatpre PON Power-On BitServicing an Operation Status Mode Event Default Power On Register States Condition Caused ByServicing Questionable Status Events Monitoring Both Phases of a Status TransitionAdding More Operation Events WAI Scpi Command CompletionDFI Discrete Fault Indicator RI Remote InhibitPage System Error Messages Error MessagesPower Supply Hardware Error Messages Calibration Error MessagesError Error String Description/Explanation/Examples Number Summary of System Error MessagesScpi Version Scpi Confirmed Commands1Scpi Approved Commands Scpi Conformance InformationNON-SCPI Commands1 Compatibility Language Compatibility LanguageParallel Polling ISET? CURR? VOUT? VoltCurr VSET?Currtrig Hold OFFVolttrig HoldSRQ FAULT?UNMASK? SRQ OFFIndex IndexOperation status group, 51 optional header Trig SOUR, 46 VOLT, 46 Volt PROT, 47 Volt Trig Japan United States Latin AmericaCanada Australia/New Zealand Europe Asia Pacific
Related manuals
Manual 142 pages 24.05 Kb Manual 67 pages 46.05 Kb

668xA, 669xA, 667xA, 664xA, 665xA specifications

Agilent Technologies has long been a pioneer in the production of high-performance electronic test and measurement instruments, particularly in the field of power sources. Among its notable offerings are the Agilent 667xA, 669xA, 665xA, 664xA, and 668xA series of power supplies. These instruments are designed to provide stable, reliable power for a variety of applications, including electronic testing, industrial processes, and research laboratories.

The Agilent 667xA series is characterized by its programmability and advanced measurement functions. These power supplies support a wide range of output voltages and currents, allowing for flexible configurations that cater to different testing needs. The built-in measurement capabilities enable users to monitor the voltage, current, and power with high precision, which is essential for ensuring optimal performance in electronic applications.

The Agilent 669xA series stands out with its high-power outputs, making it suitable for demanding applications. These power supplies deliver high voltage and current levels, making them ideal for testing high-performance devices, such as power amplifiers and motor drives. Additionally, the 669xA series includes features such as overvoltage protection and complex output sequencing to enhance the safety and reliability of the testing process.

The Agilent 665xA and 664xA series focus on delivering high accuracy and excellent regulation. These models are particularly known for their low noise operation, which is critical for sensitive applications where precision is paramount. The integrated programming capabilities allow users to automate testing sequences, thus improving efficiency in research and development settings.

The 668xA series features advanced digital signal processing that enhances the precision and stability of the output. Users benefit from features like remote sensing and monitoring, allowing feedback adjustments that maintain output accuracy despite cable losses. Furthermore, the 668xA models can integrate seamlessly with various test environments thanks to their LAN, GPIB, and USB connectivity options.

Overall, the Agilent 667xA, 669xA, 665xA, 664xA, and 668xA power supplies provide a comprehensive range of solutions for diverse electronic testing needs. With their advanced features, superb measurement capabilities, and robust performance, these instruments empower engineers and researchers to conduct their work with confidence, precision, and efficiency.