5.3.3Factory settings (Reset)

Caution!

A reset sets the instrument back to the factory settings. This can lead to an impairment of the measurement. Generally, you should perform a basic setup again following a reset.

A reset is only necessary:

if the instrument no longer functions

if the instrument must be moved from one measuring point to another

if the instrument is being de-installed /put into storage/installed

– + E

User input ("reset" (0A3)):

333 = customer parameters

555 = History

333 = reset customer parameters

This reset is recommended whenever an instrument with an unknown 'history' is to be used in an application:

The 7500 RTG is reset to the default values.

The customer specific tank map is not deleted.

A linearization is switched to "linear" although the table values are retained. The table can be reactivated in the "linearization" (04) function group.

List of functions that are affected by a reset:

linearization (041)

tank shape (002)

empty calibr. (005)

customer unit (042)

full calibr. (006)

diameter vessel (047)

pipe diameter (007)

range of mapping (052)

output on alarm (010)

pres. Map dist (054)

output on alarm (011)

offset (057)

outp. echo loss (012)

low output limit (062)

ramp %span/min (013)

fixed current (063)

delay time (014)

fixed cur. value (064)

safety distance (015)

simulation (065)

in safety dist. (016)

simulation value (066)

Tank Gauging (030)

format display (094)

auto correction (031)

distance unit (0C5)

level/ullage (040)

download mode (0C8)

555 = History Reset

After mounting and aligning the equipment, carry out a history reset before switching on the function "Auto-correction" (031) (see page 57).

The tank map can also be reset in the "cust. tank map" (055) function of the "extended calibr." (05) function group.

This reset is recommended whenever an instrument with an unknown 'history' is to be used in an application or if a faulty mapping was started:

• The tank map is deleted. The mapping must be recommenced.

39

Image 39
Contents IOM027FVAE0804 Radar Tank GaugeContrast Installation and Operations ManualBrief operating instructions Radar Tank GaugeTable of Contents Designated use Safety InstructionsInstallation, commissioning and operation Operational safetyReturn 7532 Safety InstructionsDirect voltage Device certified for use in explosion hazardous areaExplosion hazardous area Safe area non-explosion hazardous area7532 Identification IdentificationDevice designation Nameplate and Certification PlatesOrdering structure Identification Radar Tank GaugeOrdering structure 7532 RTG 7532 Output and operation CE mark, declaration of conformity Scope of deliveryCertificates and approvals Registered trademarksQuick installation guide Mounting7532 Mounting Storage Incoming acceptance, transport, storageIncoming acceptance TransportDimensions Installation ConditionsEngineering tips Measuring conditionsBehaviour if measuring range is exceeded Measuring range depending on product class for 7532 RTG 7532 Mounting Measuring rangeBlocking distance Mounting Radar Tank Gauge Blocking distanceIt is strongly recommended not to dismantle this antenna Installation instructionsInstallation in stilling well Mounting kit7532 RTG Turn housing Installation tips Installation with Varec UNI flange06-en-002 Pull off housing Mount Varec UNI flange Ansi Mounting with Sample hatch on stilling wellMounting with Pivoting element Post-installation checkWiring Quick wiring guide7532 Wiring 3 4 Wiring Radar Tank GaugeGrounding single sided on Tank Side Monitor 4590 TSM 7532 Wiring Wiring with Tank Side MonitorConnecting the measuring unit Connection to Tank Side Monitor Hart connection with two Varec RN 221 NHart connection with other supplies Post-connection check Equipotential bondingDegree of protection Overvoltage protectorOperation Quick operation guide7532 Operation Typing in numerals and text Selection and configuration in Operation menuOperation Radar Tank Gauge Selection menusIdentifying the functions General structure of the operating menuDisplay and operating elements DisplayLiquid crystal display LCD Light emitting diodes LEDs Display symbolsSymbols Meaning Calibration to regulatory standards disturbed Software reliability Custody locking switch Key assignment Function of the keysHardware lock Locking of the configuration modeLocal operation Unlock parameter 0A4Unlocking of configuration mode 555 = History Reset Factory settings ResetUser input reset 0A3 333 = reset customer parametersAlarm / Warning Display and acknowledging error messagesType of error AlarmHart communication Handheld unit DXRToF Tool operating program Operation Radar Tank Gauge Commuwin II-Operating Program ConnectionInterface Hardware Server Device list Page Commissioning Switching on the measuring deviceCommissioning Function checkCommissioning Radar Tank Gauge Basic Setup7532 SetupCommissioning Radar Tank Gauge Stilling well Basic Setup with the VUFunction group basic setup SelectionSelection Unknown Commissioning Radar Tank Gauge Function medium propertySelection Standard 7532 Commissioning Function process condFunction full calibr Commissioning Radar Tank Gauge Function empty calibrFunction dist./ meas. value Function pipe diameterDist. unknown Function range of mapping Display dist./meas.value Function history resetFunction auto-correction Mounting calibration with VUFunction group mounting calibr Function tank gaugingFunction pipe diam. corr 7532 Commissioning Display custody mode 0A9 Active posActive neg Offset Commissioning Radar Tank Gauge Dip tableDisplay Table off Table onTable off Function dip table Semi-automaticClear table ViewStore point Selection New pointPrevious point Commissioning Radar Tank Gauge Delete pointReturn Next pointSingle curve or Function plot settings 09AEnvelope curve with VU Function recording curve 09BBasic Setup /5 Basic Setup with the ToF Tool7532 Commissioning Basic Setup /5 You can also specify the pipe diameter in this display User-specific applications operation Envelope curve with the ToF ToolMounting calibration /2 Mounting calibration with the ToF Tool7532 Commissioning Mounting calibration /2 Page Replacement Maintenance7532 Maintenance Exterior cleaningPage Accessories Page Troubleshooting Troubleshooting instructions7532 Troubleshooting Troubleshooting Radar Tank Gauge Code Description Possible cause Remedy System error messagesCode Description Possible cause Remedy Error Output Application errorsLoss of echo LevelSpare parts 7532 RTG Modification nameplate Spare parts 7532 RTGSoftware changes Documentation changes Date Troubleshooting Radar Tank Gauge Housing T12Disposal Software historyTechnical data Technical data at a glance7532 Technical data According to Oiml R85 Technical dataRadar Tank GaugeOperating conditions EnvironmentProcess conditions Supplementary Documentation Human interfaceAPI American Petroleum Institute XA 081F-AAppendix Operating menu Hart Display modul, ToF Tool7532 Appendix Dome ceiling Unknown Standard Appendix Radar Tank GaugeReturn to Group Selection Operating matrix Hart / Commuwin Operating Matrix Hart / Commuwin Description of functions Function and system design Measuring principleInput Equipment architecture Appendix Radar Tank Gauge OutputStand-alone Integration into the Asset Management System 7532 Appendix On-site operationRemote operation System integration via Rackbus100 Installation and Operations Manual Custody transfer modeParticularities in approved operation Definition of terms101 Integrated on tank gauging system 102 Installation and Operations ManualIndex Radar Tank Gauge IndexIOM027FVAE0804 103 Quick wiring guide 7532 IOM027FVAE0804 105 Radar Tank Gauge Your offical representative