Agilent Technologies 8510 manual Pshort

Page 25

4.Change the class label for S11A: LABEL CLASS, S11A, ERASE TITLE.

5.Enter the label of PSHORT 1 by using the knob, the SELECT soft key and the SPACE soft key.

6.Complete the label entry procedure: TITLE DONE, LABEL DONE.

Follow a similar procedure to enter the remaining standard classes and labels as shown in the table below:

Finally, change the cal kit label as follows:

1.Press LABEL KIT, ERASE TITLE.

2.Enter the title “P BAND.”

3.Press TITLE DONE, KIT DONE (MODIFIED). The message “CAL KIT SAVED” should appear.

This completes the entire cal kit modification for front panel entry. An example of programmed modification over the GPIB bus through an exter- nal controller is shown in the Introduction To Programming section of the Operating and Service manual (Section III).

Standard

Standard

Class

class

numbers

label

 

 

 

S11B

2

PSHORT 2

S11C

3

PLOAD

S22A

1

PSHORT 1

S22B

2

PSHORT 2

S22C

3

PLOAD

FWD TRANS

4

THRU

FWD MATCH

4

THRU

REV TRANS

4

THRU

REV MATCH

4

THRU

RESPONSE

1,2,4

RESPONSE

 

 

 

25

Image 25
Contents Dis Product Information For Support Reference Only Table of contents Introduction Measurement errorsMeasurement calibration Calibration kit Class assignment Standard definitionStandard definitions table Standard class assignments Select standards Modification procedureStandard definition models Define standardsStandard type Standard numberOpen circuit capacitance C0 , C1 , C2 and C3 ΠfZ ΠfXShort circuit inductance L0 , L1, L2 and L3 ∆∅radians = 2πf ∆lengthFixed or sliding Offset delayTerminal impedance Offset Z0 Linear delay Actual delay = Fco/f2GHz C Z Offset lossLog e10 1GHzLower/minimum frequency Upper/maximum frequency ∅radians = 2π = 2πfdelayCoax or waveguide Λg = λ Co2Standard labels Assign classesStandard Classes Forward transmission match and thru S11 A,B,C and S22 A,B,CReverse transmission match and thru IsolationTRL Reflect TRL ThruTRL Line TRM ThruTRL options Standard Class labelsCalibration kit label Verify performance Enter standards/classesModification examples User modified cal kits and Agilent 8510 specificationsModeling a thru adapter Modeling an arbitrary impedance standardDisk procedure Appendix a Calibration kit entry procedureTo load calibration kits from disk into Agilent To store calibration kits from the Agilent 8510 onto a diskFront panel procedure P-band waveguide example Pshort Mm coaxial connector interface Appendix B Dimensional considerations in coaxial connectorsType-N coaxial connector interface Female type-N Page Equation Appendix C Cal coefficients modelTheir first order approximations, R is small and G=0, are Then Agilent Direct Agilent Email UpdatesAgilent Open