Agilent Technologies 8510 manual Front panel procedure P-band waveguide example

Page 24

Front panel procedure: (P-band waveguide example)

1.Prior to modifying or generating a cal kit, store one or both of the cal kits in the 8510’s non- volatile memory to a disk.

2.Select CAL menu, MORE.

3.Prepare to modify cal kit 2: press MODIFY 2.

4.To define a standard: press DEFINE STANDARD.

5.Enable standard no. 1 to be modified: press 1, X1.

6.Select standard type: SHORT.

7.Specify an offset: SPECIFY OFFSETS.

8.Enter the delay from Table 1: OFFSET DELAY, 0.0108309, ns.

9.Enter the loss from Table 1: OFFSET LOSS, 0, X1.

10.Enter the Z0 from Table 1: OFFSET Z0, 50, X1.

11.Enter the lower cutoff frequency: MINIMUM FREQUENCY, 9.487 GHz.

12.Enter the upper frequency: MAXIMUM FRE- QUENCY, 18.97 GHz.

13.Select WAVEGUIDE.

14.Prepare to label the new standard: PRIOR MENU, LABEL STANDARD, ERASE TITLE.

15.Enter PSHORT 1 by using the knob, SELECT LETTER soft key and SPACE soft key.

16.Complete the title entry by pressing TITLE DONE.

17.Complete the standard modification by press- ing STANDARD DONE (DEFINED).

Standard #1 has now been defined for a 1/8 λ P-band waveguide offset short. To define the remaining standards, refer to Table 1 and repeat steps 4 -17.

To define standard #3, a matched load, specify “fixed.”

The front panel procedure to implement the class assignments of Table 2 for the P-band waveguide cal kit are as follows:

1.Prepare to specify a class: SPECIFY CLASS.

2.Select standard class S11A.

3.Inform the 8510 to use standard no. 1 for the S11A class of calibration: l, X1, CLASS DONE (SPECIFIED).

24

Image 24
Contents Dis Product Information For Support Reference Only Table of contents Measurement errors IntroductionMeasurement calibration Calibration kit Standard definition Class assignmentStandard definitions table Standard class assignments Modification procedure Select standardsDefine standards Standard definition modelsStandard number Standard typeOpen circuit capacitance C0 , C1 , C2 and C3 ΠfX ΠfZ∆∅radians = 2πf ∆length Short circuit inductance L0 , L1, L2 and L3Offset delay Fixed or slidingTerminal impedance Linear delay Actual delay = Fco/f2 Offset Z0Offset loss GHz C ZLog e10 1GHzLower/minimum frequency ∅radians = 2π = 2πfdelay Upper/maximum frequencyCoax or waveguide Λg = λ Co2Assign classes Standard labelsStandard Classes S11 A,B,C and S22 A,B,C Forward transmission match and thruReverse transmission match and thru IsolationTRL Thru TRL ReflectTRL Line TRM ThruStandard Class labels TRL optionsCalibration kit label Enter standards/classes Verify performanceUser modified cal kits and Agilent 8510 specifications Modification examplesModeling a thru adapter Modeling an arbitrary impedance standardAppendix a Calibration kit entry procedure Disk procedureTo load calibration kits from disk into Agilent To store calibration kits from the Agilent 8510 onto a diskFront panel procedure P-band waveguide example Pshort Appendix B Dimensional considerations in coaxial connectors Mm coaxial connector interfaceType-N coaxial connector interface Female type-N Page Appendix C Cal coefficients model EquationTheir first order approximations, R is small and G=0, are Then Agilent Email Updates Agilent DirectAgilent Open