Standard definition

Standard definition is the process of mathematical- ly modeling the electrical characteristics (delay, attenuation and impedance) of each calibration standard. These electrical characteristics can be mathematically derived from the physical dimen- sions and material of each calibration standards or from its actual measured response. A standard definition table (see Table 1) lists the parameters that are used by the 8510 to specify the mathemati- cal model.

Class assignment

Class assignment is the process of organizing cali- bration standards into a format which is compati- ble with the error models used in measurement calibration. A class or group of classes correspond to the seven calibration types used in the 8510. The 17 available classes are identified later in this note (see Assign classes).

5

Image 5
Contents Dis Product Information For Support Reference Only Table of contents Measurement calibration Measurement errorsIntroduction Calibration kit Class assignment Standard definitionStandard definitions table Standard class assignments Select standards Modification procedureStandard definition models Define standardsOpen circuit capacitance C0 , C1 , C2 and C3 Standard numberStandard type ΠfZ ΠfXShort circuit inductance L0 , L1, L2 and L3 ∆∅radians = 2πf ∆lengthTerminal impedance Offset delayFixed or sliding Offset Z0 Linear delay Actual delay = Fco/f2GHz C Z Offset lossLog e10 1GHzLower/minimum frequency Upper/maximum frequency ∅radians = 2π = 2πfdelayCoax or waveguide Λg = λ Co2Standard Classes Assign classesStandard labels Forward transmission match and thru S11 A,B,C and S22 A,B,CReverse transmission match and thru IsolationTRL Reflect TRL ThruTRL Line TRM ThruCalibration kit label Standard Class labelsTRL options Verify performance Enter standards/classesModification examples User modified cal kits and Agilent 8510 specificationsModeling a thru adapter Modeling an arbitrary impedance standardDisk procedure Appendix a Calibration kit entry procedureTo load calibration kits from disk into Agilent To store calibration kits from the Agilent 8510 onto a diskFront panel procedure P-band waveguide example Pshort Type-N coaxial connector interface Appendix B Dimensional considerations in coaxial connectorsMm coaxial connector interface Female type-N Page Equation Appendix C Cal coefficients modelTheir first order approximations, R is small and G=0, are Then Agilent Open Agilent Email UpdatesAgilent Direct