Agilent Technologies 8510XF, 85107B manual Test set upgrades, Expanding measurement capabilities

Page 11

Test set upgrades

A single 8510 network analyzer can be configured to control up to four test sets. The operator can switch between test sets, without reconnections, using front panel controls. This arrangement is often referred to as the multiple test set configuration. Each test set must be equipped with Option 001 for IF switching. Other requirements may apply for RF/LO switching.

For more details, please refer to product note 8510-14, Using Multiple Test Sets with the Agilent 8510C (literature number 5967-5886E).

Agilent 8511A K01 or 08511-60008Retrofits IF switching (Option 001) to any 8510 test set for multiple test set operations.

Expanding measurement capabilities

Adding pulse capabilities (Option 008)

To perform pulse measurements, the 8510 network analyzer must be retrofitted with Option 008. The measurement system must include either the 85110L (0.045 to 2 GHz) or 85110A (2 to 20 GHz) pulsed-RF S-parameter test set and two synthesizers (high power synthesizers may apply).

The 85110L requires an 83620B synthesizer with Options 001/004/008/H80 (for RF) and an 83620B with Options 004/008/H80 (for LO)

The 85110A requires an 83622B synthesizer with Options 001/004/008 (for RF) and an 83623L with Options 004/008 (for LO)

Agilent 85111B Adds pulsed-RF measurement capability (Option 008) to the 8510C by adding new circuitry and includes on-site installation by an Customer Engineer (where available). To perform pulsed-RF measurements, the 8510 with Option 008 must be used with either the 85110L or 85110A Pulsed-RF S-parameter test set. Other requirements may apply.

Adding time domain capabilities

With the time domain option, data from transmission or reflection measure- ments are converted from the frequency domain to the time domain via the inverse Fourier transform. The time domain data are presented on the CRT display showing the measured parameter value versus time.

Agilent 85012C Adds time domain (Option 010) to an 8510C (customer installed).

11

Image 11
Contents 85107B, 45 MHz to 50 GHz in coax Agilent Upgrade Guide for the Vector Network Analyzeron-wafer configuration Product NoteTable of Contents Receiver/Detector Processor/Display Signal separationVector network analyzer VNA overview Sourceanalyzer 8510 VNA = systemSignal 8510Receiver/detector networkTable 1. Hardware revision and measurement capability Agilent 8510 family of productsAgilent 8510 network analyzer Figure 4. 8510 family of system solutions8510C Agilent 8510 Firmware upgrade kitsFigure 5. 8510 firmware upgrade kits Table 2. Firmware revision and measurement capabilityTable 3. Family of test sets Test SetsTable 5. 8360 general purpose synthesizers SourcesTable 4. 8510-dedicated sources 85103E Upgrade considerationsUpgrading individual components Upgrading theFigure 7. 8510 to 8530 upgrade paths 8510C8530A 8530A microwave receiver and 8510CAdding time domain capabilities Test set upgradesExpanding measurement capabilities Adding pulse capabilities OptionFigure 8. 8510 family of system solutions + ORUpgrading systems 3. What source do I need? System upgrades4. What measurement accessories do I need? 2. What test set do I need?Upgrades for 85107B, 85109C 85107B 45 MHz to 50 GHz current system Examples of upgrade pathsMeasurement accessories Upgrade to 85106D waveguide system to operate in V-bandCalibration kits Device under test Verification kitsTest port cables and adapters Configuration A Network analyzer test setFor 8517B Test sets 2.4 mm rugged test port connectors Configuration B Network analyzer test set Cable set Device under testFor 8514B/8515A/85110A test sets 3.5 mm rugged test port connectors For 85110L Test sets 7 mm rugged test port connectorsProduct notes Related literaturePub. number Product overviewsOur Promise Agilent Technologies’ Test and Measurement Support, ServicesFor more information on the 8510 VNA visit and Assistance