Teledyne T200U-NOy manual Initial Operation, Final Test and Calibration Values

Page 17

Model T200U-NOy Option Manual Addendum

3.3Initial Operation

1.After confirming proper supply voltage, turn on the power to the analyzer and the 501Y bypass pump, and plug in the external Pump. If you are unfamiliar with the T200U-NOy analyzer, we recommend that you read the overview in Section 3 of its operation manual before proceeding.

The front panel displays should immediately light as the instruments start up.

The 501Y requires about 30 minutes for the external Converter to come up to operating temperature.

The T200U-NOy requires about 30 minutes for the ozone generator to start up. During that time the instrument will not respond to span gas.

After 30 min, the display on the 501Y front panel should read 315oC, indicating that the external Converter is at operating temperature.

Follow the instructions in the T200 Manual and the T200U Addendum to confirm proper operation of the analyzer.

2.Proceed to Section 4 of this manual to perform a Zero/Span check.

Table 2 Final Test and Calibration Values

 

Test Values

 

 

Observed

Units

Nominal

 

 

 

 

 

 

Value

 

 

Range

 

 

CONVERTER TEMP

 

 

 

oC

315

7

 

 

BYPASS

FLOW

NO

 

CC/MIN

800

 80

 

 

CHANNEL

 

 

 

 

 

 

 

 

 

BYPASS

FLOW

NOy

 

CC/MIN

800

± 80

 

 

CHANNEL

 

 

 

 

 

 

 

 

Converter Serial # __________________________

 

 

 

 

Date _______________________

 

 

Technician _____________________

07303C DCN6646

15

Image 17
Contents T200U-NOy Analyzer Page Safety Messages Consignes DE Sécurité Warranty This page intentionally left blank Table of Contents List of Figures Introduction 07303C DCN6646 Specifications SpecificationsModel T200EU-NOy Option Manual Addendum 2.0 Specifications Unpacking and Inspection Getting StartedElectrical and Pneumatic Connections Model T200U-NOy Option Manual Addendum Rear Panel Pneumatic Connections Final Test and Calibration Values Initial OperationThis page intentionally left blank Principle of Operation T200U-NOY External ConverterThis page intentionally left blank Calibration and CAL Checks Manual Zero Calibration Procedure Calibration or Cal Check ProcedureSpan Calibration Procedure This page intentionally left blank Maintenance Schedule MaintenanceReplacing the Sample Particulate Filters Replacing the Particulate Filter Replacing the Converter Checking Analyzer Flow RateInspecting Pneumatic Lines NOy Converter Guts Assembly LocationPneumatics Location of O3 Generator Air Inlet Checking for LeaksPressure must be less than 15 PSI Operation Verification Diagnostic Techniques TROUBLESHOOTING, Adjustments3 501Y Bypass Pump Diagnostic Procedures Setting the Converter TemperatureLOC Lock AH Alarm Electrical Diagram This page intentionally left blank Partnumber Description T200U-NOY Spare Parts ListPCA, PMT PREAMP, VR FILTER, SS KB This page intentionally left blank

T200U-NOy specifications

The Teledyne T200U-NOy is an advanced gas analyzer specifically designed to measure nitrogen oxides (NOx) concentrations in a variety of applications, including industrial emissions monitoring, environmental assessments, and combustion control. It builds upon Teledyne's legacy of producing high-quality analytical instruments, making it a reliable choice for professionals in the field.

One of the main features of the T200U-NOy is its ability to measure multiple nitrogen oxides, including nitrogen dioxide (NO2), nitric oxide (NO), and nitrous oxide (N2O), which are critical for understanding and managing air quality and combustion efficiency. The instrument offers high sensitivity and accuracy, allowing for real-time monitoring of NOx levels with a detection limit that suits stringent regulatory requirements.

The T200U-NOy utilizes advanced chemiluminescence technology for NO measurement, which is widely recognized for its reliability and precision. This method involves the reaction of NO with ozone (O3) to produce excited NO2, which subsequently releases light as it returns to a lower energy state. The intensity of this light is measured and directly correlates to the concentration of NO in the sample, providing highly accurate readings.

In addition to its core measurement capabilities, the T200U-NOy is equipped with a variety of features that enhance its functionality. Its user-friendly interface allows for easy configuration and data retrieval, making it accessible to a wide range of users, from seasoned professionals to those new to gas analysis. The device supports multiple communication protocols, including analog and digital outputs, facilitating integration into existing monitoring systems.

The T200U-NOy is constructed with robust materials designed to withstand harsh industrial environments, ensuring longevity and reliability. It also includes features for easy maintenance, such as user-replaceable parts, which can significantly reduce downtime.

Overall, the Teledyne T200U-NOy stands out in the field of gas analyzers, combining cutting-edge measurement technologies with user-centric features and robust construction. Its comprehensive capabilities make it an essential tool for accurately monitoring nitrogen oxides, ultimately contributing to better environmental protection and compliance with air quality regulations.