7.0 TROUBLESHOOTING, ADJUSTMENTS

CAUTION!

The operations outlined in this section must be performed by qualified maintenance personnel only!

This section of the manual contains information on diagnosing and repairing instrument performance problems. It provides troubleshooting procedures that address problems to the board level.

As a guide to troubleshooting, think of the T200U-NOy as two systems: pneumatics and electronics.

Pneumatics - Over 50% of all instrument problems are traced to pneumatic leaks. Suspect a leak first.

Electronics – external Converter temp control board or DC power supply for the converter temperature readout.

Another trouble-shooting tip has to do with incorrect span gas concentration:

This could come either from the calibrator itself or from incorrectly entering the expected span gas concentration in the instrument.

If the instrument does not respond to span gas, check troubleshooting section in the T200 manual.

The above should get you started in diagnosing and repairing the most common faults. If these reasons have been eliminated and there are still problems with the unit, the next thing to do is a Factory Calibration covered in the T200 manual. If difficulties persist, contact our service department. The 800 telephone number is on the cover page of this manual.

7.1 Operation Verification – Diagnostic Techniques

This section provides guidance on diagnosing some possible problems.

7.1.1 Pneumatic System

Refer to Figure 5 for a diagram of this system’s pneumatics, to be used in conjunction with diagnostics and troubleshooting.

7.1.2 Leak Check

Refer to Section 6.6 for the leak check procedure.

07303C DCN6646

31

Image 33
Contents T200U-NOy Analyzer Page Safety Messages Consignes DE Sécurité Warranty This page intentionally left blank Table of Contents List of Figures Introduction 07303C DCN6646 Specifications SpecificationsModel T200EU-NOy Option Manual Addendum 2.0 Specifications Unpacking and Inspection Getting StartedElectrical and Pneumatic Connections Model T200U-NOy Option Manual Addendum Rear Panel Pneumatic Connections Final Test and Calibration Values Initial OperationThis page intentionally left blank Principle of Operation T200U-NOY External ConverterThis page intentionally left blank Calibration and CAL Checks Manual Zero Calibration Procedure Calibration or Cal Check ProcedureSpan Calibration Procedure This page intentionally left blank Maintenance Schedule MaintenanceReplacing the Sample Particulate Filters Replacing the Particulate Filter Replacing the Converter Checking Analyzer Flow RateInspecting Pneumatic Lines NOy Converter Guts Assembly LocationPneumatics Location of O3 Generator Air Inlet Checking for LeaksPressure must be less than 15 PSI Operation Verification Diagnostic Techniques TROUBLESHOOTING, Adjustments3 501Y Bypass Pump Diagnostic Procedures Setting the Converter TemperatureLOC Lock AH Alarm Electrical Diagram This page intentionally left blank Partnumber Description T200U-NOY Spare Parts ListPCA, PMT PREAMP, VR FILTER, SS KB This page intentionally left blank

T200U-NOy specifications

The Teledyne T200U-NOy is an advanced gas analyzer specifically designed to measure nitrogen oxides (NOx) concentrations in a variety of applications, including industrial emissions monitoring, environmental assessments, and combustion control. It builds upon Teledyne's legacy of producing high-quality analytical instruments, making it a reliable choice for professionals in the field.

One of the main features of the T200U-NOy is its ability to measure multiple nitrogen oxides, including nitrogen dioxide (NO2), nitric oxide (NO), and nitrous oxide (N2O), which are critical for understanding and managing air quality and combustion efficiency. The instrument offers high sensitivity and accuracy, allowing for real-time monitoring of NOx levels with a detection limit that suits stringent regulatory requirements.

The T200U-NOy utilizes advanced chemiluminescence technology for NO measurement, which is widely recognized for its reliability and precision. This method involves the reaction of NO with ozone (O3) to produce excited NO2, which subsequently releases light as it returns to a lower energy state. The intensity of this light is measured and directly correlates to the concentration of NO in the sample, providing highly accurate readings.

In addition to its core measurement capabilities, the T200U-NOy is equipped with a variety of features that enhance its functionality. Its user-friendly interface allows for easy configuration and data retrieval, making it accessible to a wide range of users, from seasoned professionals to those new to gas analysis. The device supports multiple communication protocols, including analog and digital outputs, facilitating integration into existing monitoring systems.

The T200U-NOy is constructed with robust materials designed to withstand harsh industrial environments, ensuring longevity and reliability. It also includes features for easy maintenance, such as user-replaceable parts, which can significantly reduce downtime.

Overall, the Teledyne T200U-NOy stands out in the field of gas analyzers, combining cutting-edge measurement technologies with user-centric features and robust construction. Its comprehensive capabilities make it an essential tool for accurately monitoring nitrogen oxides, ultimately contributing to better environmental protection and compliance with air quality regulations.