Model GFC7000T Carbon Dioxide Analyzer
1. OVERVIEW
The Model GFC 7000T is a close derivative of the Model GFC 7000TA CO2 Analyzer. The main differences are as follows.
The optical bench is longer (14 m), making the instrument more sensitive at low CO2 levels. The instrument has a
The rear panel is different. There is an extra pneumatic port for the exhaust of the purge gas to the GFC wheel. Also, there are relay alarm outputs available on the rear panel.
Refer to the main manual (i.e. GFC 7000TA) that is shipped with your instrument for details of the instrument’s components and operation. Any differences between this instrument and what is described in the GFC 7000TA, will be described in this Addendum to the manual.
2. SPECIFICATIONS AND AGENCY APPROVALS
Table
| Table |
|
| |
|
|
|
|
|
| PARAMETER | SPECIFICATION |
|
|
| Ranges | User selectable to any full scale range from |
|
|
|
|
|
|
|
| Measurement Units | ppb, ppm, µg/m3, mg/m3 (user selectable) |
|
|
| Zero Noise | < 0.1 ppm RMS |
|
|
|
|
|
|
|
| Span Noise | < 1% of reading RMS |
|
|
|
|
|
|
|
| Lower Detectable Limit | < 0.2 ppm |
|
|
| Zero Drift (24 hours) | < 0.25 ppm |
|
|
|
|
|
|
|
| Span Drift (24 hours) | < 0.5% of reading |
|
|
|
|
|
|
|
| Linearity | 1% of full scale |
|
|
|
|
|
|
|
| Precision | 0.5% reading |
|
|
|
|
|
|
|
| Lag Time | 10 sec |
|
|
| Rise/Fall Time | <60 sec to 95% |
|
|
|
|
|
|
|
| Sample Flow Rate | 800 cm3/min. ± 10% |
|
|
| Temperature Range | 5 - 40°C operating |
|
|
|
|
|
|
|
| Humidity Range |
|
| |
|
|
|
|
|
| Temp Coefficient | < 0.05 % of reading per °C |
|
|
| Voltage Coefficient | < 0.05 % of reading per V |
|
|
|
|
|
|
|
| Dimensions (HxWxD) | 7" x 17" x 23.5" (178 mm x 432 mm x 597 mm) |
|
|
|
|
|
|
|
| Weight | 40 lb (18.1 kg) |
|
|
|
|
|
|
|
| AC Power | 100 – 120V 50/60 Hz, |
|
|
|
|
|
| |
|
|
|
|
|
|
| Teledyne Analytical Instruments | 1 |