5.RELAY ALARM OUTPUTS (STANDARD CONFIGURATION)

Model GFC7000T Carbon Dioxide Analyzer

EXAMPLE: If the application is to measure between 0 ppm and 50 ppm, an appropriate Span Gas would be 40 ppm. If the application is to measure between 0 ppm and 100 ppm, an appropriate Span Gas would be 80 ppm.

Zero Air is similar in chemical composition to the earths atmosphere but scrubbed of all components that might affect the analyzer’s readings. In the case of CO2 measurements this means CO2 less than 0.1 ppm of CO2 and Water Vapor.

Zero Air can be purchased in pressurized canisters or created using a Teledyne Instruments Model 701 Zero Air Generator in combination with a canister of indicating soda-lime.

4.Attach an exhaust line to the exhaust outlet port.

The exhaust from the analyzer and vent lines should be vented to atmospheric pressure using maximum of 10 meters of 1/4” PTFE tubing.

5.Attach a source of dried air scrubbed of CO2 to the purge inlet port

The source of purge gas should be at least 10 psig and capable of maintaining a flow of at least 1 liter/min.

Purge source gas pressure should not exceed 30 psig. The purge exhaust is only used on the GFC 7000TU model.

6.Once the appropriate pneumatic connections have been made, check all pneumatic fittings for leaks.

5.RELAY ALARM OUTPUTS (STANDARD CONFIGURATION)

There are 4 relay alarm outputs (AL1-AL4) on the rear panel.

AL1 is for system okay,

AL2 is for concentration limit 1 exceeded and

AL3 is for concentration limit 2 exceeded.

AL4 is not used

The relay alarm output AL1 is enabled all the time, whereas the AL2 and AL3 can be enabled/disabled by going into the diagnostics menu (with 929 password), then factory options, then turning conc. Alarm Relays ON or OFF.

When the concentration alarm relays are enabled, concentration alarms status is no longer available through the “Status output” pins. So, either the alarm relays or the status bits could be used to monitor the concentration alarms but not both at the same time. The default factory setting is to enable the concentration alarm relays.

The AL1 relay is energized when the system is okay and de-energized when the system has a fault. The AL2 and AL3 relays energize when the corresponding concentration limits are exceeded.

Teledyne Analytical Instruments

6

Image 12
Contents Teledyne Analytical Instruments Model GFC 7000T CO2 AnalyzerManual Addendum Addendum to GFC 7000TA Manual, PNModel GFC7000T Carbon Dioxide Analyzer WarrantyCopyright 2013 Teledyne Analytical Instruments Important NoticeABOUT THIS ADDENDUM This page intentionally left blank TABLE OF CONTENTS Model GFC7000T Carbon Dioxide Analyzer 2. SPECIFICATIONS AND AGENCY APPROVALS 1. OVERVIEWTable 2-1. GFC 7000T Specifications and Agency Approvals SPECIFICATION3. ELECTRICAL CONNECTIONS 4.1. BASIC AND ZERO/SPAN VALVE OPTION CONFIGURATIONS 4. PNEUMATIC CONNECTIONSFigure 4-1. Pneumatic Connections for Pressurized Sample Delivery w/o Internal Pump 4.2. MAKING THE PNEUMATIC CONNECTIONS 5. RELAY ALARM OUTPUTS STANDARD CONFIGURATION 6. RELAY ALARM OUTPUTS AIR PRODUCTS CONFIGURATION 7. CONTROL INPUTS AIR PRODUCTS CONFIGURATION ONLY