Model GFC7000T Carbon Dioxide Analyzer
4. PNEUMATIC CONNECTIONS
This section provides information on the basic pneumatic configuration and the zero/span valve option pneumatic configuration. First are diagrams to illustrate the configurations, followed by a description of the pneumatic ports, and then
4.1.BASIC AND ZERO/SPAN VALVE OPTION CONFIGURATIONS
CAUTION
Sample and calibration gases should only come into contact with PTFE
(Teflon), FEP, glass, stainless steel or brass.
NOTE
To prevent dust from entering the gas flow channels, your analyzer was shipped with small plugs inserted into each of the pneumatic fittings on the back panel. Remove these dust plugs and store for future use before proceeding.
Figure
Table
Note that the flowmeter could also be configured upstream of the instrument. Since most flowmeters are calibrated at ambient pressure, ensure that the flow going through the flow meter is at ambient pressure, when it is placed upstream.
Teledyne Analytical Instruments | 3 |