Model GFC7000T Carbon Dioxide Analyzer
Table
Rear Panel Label | Function | |
| Connect a gas line from the source of sample gas here. | |
SAMPLE | Calibration gasses are also inlet here on units without | |
| zero/span valve option installed. | |
|
| |
EXHAUST | Connect an exhaust gas line of not more than 10 meters | |
long here. | ||
| ||
|
| |
PRESSURE SPAN | On units with zero/span valve option installed, connect a | |
| gas line to the source of calibrated span gas here. | |
VENT SPAN | Not used | |
|
| |
ZERO | On units with zero/span valve option installed, attach a | |
gas line to the source of zero air here. | ||
| ||
|
| |
| This inlet supplies purge air to the GFC wheel housing | |
TO PURGE | Connect a source of dried air that has been scrubbed of | |
| CO2. | |
| This exhausts purge air to the GFC wheel housing. | |
FROM PURGE | Connect an exhaust gas line of not more than 10 meters | |
long here. It is only used on the GFC 7000TU instrument. | ||
| ||
|
|
4.2.MAKING THE PNEUMATIC CONNECTIONS
CAUTION
Venting should be outside the shelter or immediate area surrounding the instrument.
1.Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more than 2 meters long.
2.Attach sources of zero air and span gas
3.Span Gas is a gas specifically mixed to match the chemical composition of the type of gas being measured at near full scale of the desired measurement range.
When CO2 measurements are to be made with the Teledyne Instruments Model GFC 7000T Analyzer it is recommended that you use a gas calibrated to have a CO2 content equaling 80% of the range of compositions being measured.
Teledyne Analytical Instruments | 5 |