WARNING: Use only probes provided by GF Health Products, Inc. intended to be used with the Grafco Pocket Doppler.
WARNING: Do not stretch the probe cable further than 6.5 feet (2 meters).
WARNING: Electromagnetic Interference: Before operation, ensure that the environment in which the device is operated is not subject to any source of strong electromagnetic interference, such as radio transmitters, mobile telephones, etc.
WARNING: The following safety checks/tests must be performed at least once every two years or
as specified in your facility's test and inspection protocol, by a qualified person with adequate training, knowledge, and practical experience:
•Inspect the equipment for mechanical and functional damage.
•Ensure the safety labels are legible.
•Verify that the device functions properly as described in this manual.
•Test according to the pregnant woman's leakage current (IEC
The leakage current should never exceed the specified limit. Record the data and store with the Grafco Pocket Doppler. If the device is not functioning properly or fails any of the above tests, contact GF Health Products, Inc. Technical Support at the phone number on the back cover, or your GF Health Products, Inc. authorized distributor.
| 7 |