WARNING: Use only probes provided by GF Health Products, Inc. intended to be used with the Grafco Pocket Doppler.

WARNING: Do not stretch the probe cable further than 6.5 feet (2 meters).

WARNING: Electromagnetic Interference: Before operation, ensure that the environment in which the device is operated is not subject to any source of strong electromagnetic interference, such as radio transmitters, mobile telephones, etc.

WARNING: The following safety checks/tests must be performed at least once every two years or

as specified in your facility's test and inspection protocol, by a qualified person with adequate training, knowledge, and practical experience:

Inspect the equipment for mechanical and functional damage.

Ensure the safety labels are legible.

Verify that the device functions properly as described in this manual.

Test according to the pregnant woman's leakage current (IEC 60601-1/1988: Limit: 100 uA (B)).

The leakage current should never exceed the specified limit. Record the data and store with the Grafco Pocket Doppler. If the device is not functioning properly or fails any of the above tests, contact GF Health Products, Inc. Technical Support at the phone number on the back cover, or your GF Health Products, Inc. authorized distributor.

14423-INS-LAB-RevE13

7

Image 7
Contents INS-LAB-RevE13 Contents IntroduCtIon InCluded featuresIntended use of thIs devICe Product description OptIonal aCCessorIesSafety preCautIons Safety InformatIonSafety guidance INS-LAB-RevE13 INS-LAB-RevE13 INS-LAB-RevE13 Symbols Gf health produCts, InC. freIght polICy HandlIngUnpaCkIng Inspection StorageAppearanCe and features Ultrasound poCket doppler and aCCessorIesButton PowerSocket Jack Definition Rear panel Feature Function BatteryProbe frame Top panel FeatureSocket Name Picture Start/stop button ReC/play button Feature Function Volume ControlMode button Left panelPoCket doppler Waterproof probes BatteryInstallIng or replaCIng the battery SetupINS-LAB-RevE13 Placing the probe in the pocket doppler OperatIonProbe operatIon Removing the probe from the pocket dopplerSwapping probes TurnIng on the poCket dopplerFh monitoring with 2.0 mhz Waterproof probe Fetal heart fh monItorIngVasCular monItorIng WIth 4.0 or 8.0 mhz probe optIonal TurnIng off the poCket doppler ReplaCIng the batteryMaIntenanCe MaIntenanCe and CleanIngCleaning the pocket doppler CleanIngCleaning the pocket doppler probe Disinfecting the pocket doppler probe DIsInfeCtIonSafety SpeCIfICatIonsPhysical Power BatteryFhr Ultrasound LImIted Warranty INS-LAB-RevE13 GF Health Products, Inc. Graham-Field EC declaration of Conformity AppendIx a, eC deClaratIon of ConformItyGuIdanCe and manufaCturer’s deClaratIon AppendIx b, emC InformatIonRF IEC DistanceRated Separation distance according to Recommended separation distancesFrequency of transmitter m INS-LAB-RevE13 Overall sensitivity 2 mhz probe AppendIx C, overall sensItIvIty=Ad+B+C Index INS-LAB-RevE13