4.4SIGNAL INPUTS
RATIO: The reference input allows an external voltage to be used as the reference source for conversion. In this mode, the meter reads the ratio of the signal voltage to the reference voltage rather than the true value of the input.
Reading in Counts = | Signal Voltage | x 10000 |
Reference Voltage |
On the 20 V and 200 V ranges, the signal voltage must be scaled by 1/10 and 1/100, respectively. For all ranges, the standard reference input impedance ratio is 80 ohm. For 100 Mohm, open solder switch F on the main board (Figure 5.3). The reference voltage must be between the limits specified, +0.5 to +2.0 V, and must be positive with respect to analog ground.
HOLD: When high (or open), the A/D will
BLANKING: The display may be blanked by grounding the BLANKING input. The blanking input must be open for normal display operation. The polarity sign and decimal points are not blanked, but they will flash if the displayed reading exceeds ±19999 counts.
8