Sigma QUick Infrared Camera manual Qcdcom Setup and Operation, Fp qrc

Page 7

October 30, 1997

5

3.2QCDCOM Setup and Operation

The qcdcom program should be run from the directory where you want your data to be written, though this can be changed from within qcdcom using the cd command. So execute the following commands:

%cd /scr/data (or wherever your data should go)

%qcdcom

When qcdcom is started, it initializes communications with the parallel interface, allocate memory into which to read IR array images, and read enough from the DSP memory to get an idea of the status of the DSP code. Qcdcom also initializes itself for readout size and binning parameters from the numbers it reads from the DSP. Since qcdcom runs separately from the DSP computer and saves its own parameters, qcdcom can be terminated and then restarted without disturbing the DSP.

At startup qcdcom prints a short message identifying the program, and display the filters installed and the current filter and pupil position. It will also try establishing communication with the camera, and will give a warning if it fails. If the camera has not been initialized after powerup, or if for some other reason communications have failed, the DSP code will have to be downloaded to the camera. This is done with the df command. You should do this at the beginning of your observing run to make sure the correct code is running. The command to download the current standard DSP code is >the following:

df quirc

Occasionally the download will fail if the camera is in some undetermined state, and a reset will need to be done. The DSP reset is a red button on the white controller box mounted on the dewar. Usually pressing this once and waiting several seconds before retrying the download will work. Once the download has completed, the sync command can be used to see if the camera communications are OK. You can use the sync command at any time to verify that qcdcom and the DSP computer are talking. Once the main DSP program has been downloaded with the df command, the utility board >program must be downloaded using the following command:

du util

This loads the code that performs the filter wheel and pupil motions.

Next you should set the way files will be named on the disk. Files are named using a prefix with the frame number appended, thus you should set the prefix using the fp command and reset the running file number using the fn command (typically to 1 when you start up). For example, to have files named qrc.001, qrc.002, etc.:

>fp qrc.

>fn 1

Note you must specify a trailing dot in the file prefix if you want one.

You will also want to set the exposure parameters using auto at this time. Since you may inherit parameters from the previous observer, you should make sure they are set to the default values, which are correct for most observing. One easy way to do this is to use the init command described

Image 7
Contents Observer’s Guide for Quirc Page Near-Infrared Observing Techniques System OverviewDetector Linearity, Saturation, Read Noise, Dark Current Dewar Temperature Setup and Observing with QuircCamera Sensitivity Telconfig06quirc Workstation setupFp qrc Qcdcom Setup and OperationTaking exposures Writing Scripts and Taking Mosaics Filters and Focusing8QUIRC User Guide FocusingRunning VF to display data Cold RestartTaking Dark Frames Image display options Setting upQuick VF tips DSP initialization commands Qcdcom commandsCommands for setting up and taking exposures October30 Filter hnum commandiSet filter Go hnumiStart exposure Status shows the current parameters Diagnostic commands Program counts the errors