To change the AC operating voltage selection:

This procedure should be performed only by qualified technical personnel. If you are not qualified, please consult TRUE Systems or your dealer for assistance.

2

1

1)Using an appropriate AC voltmeter, measure the AC line voltage at the installation site.

2)Make sure the AC power cord is disconnected from the PSU-P8i Power Supply Unit.

3)Using a small screwdriver or similar tool, release the latch on the Fuse

Drawer 1 and pull it (and the fuses) out of the AC inlet module.

4)Using a small screwdriver or similar tool, remove the Voltage Selector 2 .

5)Rotate the Voltage Selector to the appropriate position for the AC voltage measured in Step 1) above. See Table below for proper Voltage Selector setting. Select the setting that places your measured voltage nearest the middle of the Operating Voltage Range.

Voltage Selector

Operating Voltage Range

Fuses (2)____

100V

90-110VAC 50/60Hz

T 500mA L, 250V

 

120V

110-132VAC 50/60Hz

T 500mA L, 250V

220V

201-240VAC 50/60Hz

T 315mA L, 250V

240V

220-264VAC 50/60Hz

T 315mA L, 250V

For other operating voltage options, consult TRUE Systems.

CAUTION: To provide continued protection against fire, replacement fuses should be only of the types listed above. Note that two (2) fuses are necessary and that both fuses

9

Image 9
Contents True Important Information Safety and Operating PrecautionsOwner’s Record Declaration of Conformity -CEContents Overview IntroductionIf you’re using the DI’s for electric instrument input If you’re using the M/S DecoderSet Peak Reference selector to +26 Unpacking and Inspection About the Level and Peak-Hold IndicatorsPSU-P8i Precision 3 4 5 6 + S + S + S + S + S + S + S + Word About Cables True E C is IO N More About M-S Decoding Optional Level Indicator Modes Design Approach Operation Symptom SolutionRegistration and Warranty Service Information