ResMed POWER STATION II, RPS II manual Technical specifications

Page 11

Technical specifications

Battery technology

Lithium ion

Capacity

< 100 Wh (97 Wh)

UN classification

UN3481 (lithium-ion batteries packed with equipment)

Power supply

Input range 100–240V, 50–60Hz, 1.0-1.5A

 

Nominal for aircraft use 110V, 400Hz

Battery output voltage

(24V / 26V) ± 0.5V, 90W continuous

Power supply unit output

3.75 A

current (rated)

 

Battery output current (rated)

3.75 A / 3.46 A

Standby current

< 100 μA

Protection

Overcharge, overdischarge, excess current, short circuit,

 

high temperature

Dimensions (L x W x H)

230 mm x 126 mm x 26 mm

Battery weight

0.9 kg

System weight

2.3 kg

Minimum life cycle

≥ 500 cycles at 23°C to 60% capacity

Operating conditions

 

Charging

5°C to 40°C; 5–85% maximum humidity

Discharging

-5°C to 40°C; 5–85% maximum humidity

Transport/storage conditions

-20°C to +45°C; 5–85% maximum humidity

Operating/storage air pressure

600 hPa to 1100 hPa

Recharge time

< 4 hrs at full level

Electromagnetic compatibility

Product complies with all applicable electromagnetic

 

compatibility requirements (EMC) according to IEC60601-

 

1-2, for residential, commercial, and light industry

 

environments. Information regarding the electromagnetic

 

emissions and immunity of these ResMed devices can

 

be found on www.resmed.com, on the Products page

 

under Service and Support. Click on the PDF file for your

 

language.

IEC 60601-1 classification

Class II (double insulation) and/or Internally Powered

 

Equipment, IP21 (IP20 while charging), Continuous

 

Operation (from mains), Limited Operation (from

 

battery), Equipment not suitable for use in the presence

 

of a flammable anaesthetic mixture with air or with

 

oxygen or nitrous oxide.

Battery run time

> 8 hrs at average device settings1

 

For more information, see the Battery/Device

 

Compatibility List on www.resmed.com.

1Using 15 cm H2O (IPAP), 5 cm H2O (EPAP), and 15 BPM (Respiratory Rate). Not applicable to devices when using heated humidification and heated tubing.

Note: The manufacturer reserves the right to change these specifications without notice.

6

Image 11
Contents Power Station Power Station 装置样品只是图解 / 裝置只是樣品圖解 DC output voltage & DC cables Svenska Dansk Norsk Suomi Ελληνικά 简体中文 繁體中文 Control panel Compatible devices and accessoriesIntended use RPS II at a glanceCharging the battery Setting upIndicators Servicing Using humidifiersCleaning and maintenance StoringYour device user guide, or the Battery/Device TroubleshootingTechnical specifications General warnings and cautions SymbolsRPS Limited warranty Page Respiratory Care Solutions