Omron HCG-801 instruction manual SD memory error Trying to erase ECG data

Page 49

HCG-801-E.book Page 49 Tuesday, December 13, 2011 5:35 PM

 

 

 

 

 

 

 

 

 

 

10. Care and Maintenance

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Error #

Message

 

 

Cause

Remedy

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SD memory error

An error occurred while

 

 

 

 

 

21

trying to read ECG data

Format the SD memory card.

 

 

 

 

Cannot read data.

stored on the SD memory

 

 

 

 

 

Refer to p. 19.

 

 

 

 

 

 

card.

*Copy any important data to a PC

 

 

 

 

 

 

An error occurred while

before formatting the SD memory

 

 

 

 

22

SD memory error

trying to erase ECG data

card as all data on the card will be

 

 

 

 

Cannot erase data.

stored on the SD memory

erased.

 

 

 

 

 

Use a new/different SD memory

 

 

 

 

 

 

card.

 

 

 

 

 

 

card. (Always turn off the unit before

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SD memory error

An error occurred while

removing or inserting SD memory

 

 

 

 

23

trying to save ECG data on

cards.)

 

 

 

 

Cannot write data.

 

 

 

 

 

the SD memory card.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

An error occurred when

Use a new/different SD memory

 

 

 

 

 

Error formatting SD

card. (Always turn off the unit

 

 

 

 

24

formatting the SD memory

 

 

 

 

memory.

before removing or inserting SD

 

 

 

 

 

card.

 

 

 

 

 

 

memory cards.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Format the SD memory card.

 

 

 

 

 

 

 

 

 

 

Refer to p. 19.

 

 

 

 

 

 

The SD memory card may

*Copy any important data to a PC

 

 

 

 

 

SD memory error.

before formatting the SD memory

 

 

 

 

25

be damaged.

card as all data on the card will be

 

 

 

 

SD memory unknown.

 

 

 

 

 

 

 

 

 

erased.

 

 

 

 

 

 

 

 

 

 

Use a new/different SD memory

 

 

 

 

 

 

 

 

 

 

card. (Always turn off the unit before

 

 

 

 

 

 

 

 

 

 

removing or inserting SD memory

 

 

 

 

 

 

 

 

 

 

cards.)

 

 

 

 

 

SD memory full.

The SD memory card is full

 

 

 

 

 

26

and there is not enough

 

 

 

 

 

Please delete

Either delete unnecessary data

 

 

 

 

space to save new ECG

 

 

 

 

 

unnecessary data.

data.

or use a new SD memory card.

 

 

 

 

 

 

 

 

 

 

(Always turn off the unit before

 

 

 

 

 

The number of data in

The SD memory card is full

 

 

 

 

 

removing or inserting SD

 

 

 

 

 

SD is over the limit.

and there is not enough

 

 

 

 

27

memory cards.)

 

 

 

 

Please delete

space to save new ECG

 

 

 

 

 

 

 

 

 

 

 

unnecessary data.

data.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Turn the unit off, then on again.

 

 

 

 

 

 

The unit may have been

If this error still occurs, consult

 

 

 

 

91

Device error.

damaged or is otherwise

your Omron Customer Service

 

 

 

 

 

 

unable to operate normally.

Department (Tel. 0870-

 

 

 

 

 

 

 

 

 

 

7502771).

 

 

 

 

 

 

 

 

 

 

49

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EN

Image 49
Contents Portable, cordless, single-channel ECG Monitor Model HCG-801Contents Intended Use IntroductionThis unit has no function to detect pacing spikes General Main Unit Know Your UnitReplacements Parts Display Settings SET menu screenMenus Menu screenECG Graph Inserting and Replacing Batteries Battery life Battery life and replacementHCG-801-E.book Page 12 Tuesday, December 13, 2011 535 PM Setting the Date and Time Unit beeps, then the power is turned off Inserting and Ejecting SD Memory Card Using SD Memory CardInserting the SD memory card Never insert an SD memory card while the unit is turned onWrite Protection Locking the SD Memory Card Close the SD memory card slot cover To eject the SD memory card Only eject an SD memory card when the unit is turned offFormatting an SD Memory Card Never remove an SD memory card while it is being formatted Do not move during measurement Important Notes on Taking a MeasurementTaking a Measurement Press closely to your chest Female About 30 seconds HCG-801-E.book Page 25 Tuesday, December 13, 2011 535 PM Common Causes of Inaccurate Measurement Results There are 13 messages regarding the measurement result Measurement ResultRecommendation on measurement result Displaying ECG Data Displaying ECG Data Basic Operation Adjusting the SettingsAdjusting the Settings Date and Time VolumeBrightness Contrast Language Date Format Messages About the Overwrite message OverwriteErasing All ECG Data Managing ECG Data Erasing ECG DataMessage Erase all data in SD memory? is displayed Erasing Selected ECG Data HCG-801-E.book Page 41 Tuesday, December 13, 2011 535 PM Copying Data to an SD Memory Card Press the OK button ECG data is copied to the SD memory card Copying data from the SD memory card to a PC Copying data from the SD memory card to a PCCopying data from a PC to the SD memory card Lists and Explanations of Messages Care and MaintenanceMessage Cause Remedy Menu Error MessagesSD memory error Trying to erase ECG data Problem Cause Solution TroubleshootingMaintenance and Storage Internally Powered Type BF Technical DataTechnical Data General Information About the Heart and ECG Measurements What is Ischemia? What is Arrhythmia?About the ECG Waveform Omron Healthcare Europe B.V
Related manuals
Manual 128 pages 10.78 Kb

HCG-801 specifications

The Omron HCG-801 is a sophisticated and advanced body composition monitor designed to provide accurate and comprehensive health assessments for users seeking to manage their weight and overall fitness. This innovative device integrates cutting-edge technology with user-friendly features to deliver reliable data on body composition metrics.

One of the standout features of the Omron HCG-801 is its ability to measure body weight, body fat percentage, visceral fat, muscle mass, bone mass, and metabolic age. By utilizing a bioelectrical impedance analysis (BIA) method, the device sends a low-level electrical current through the body to estimate the composition of tissues. This technology allows users to gain insight into their body’s health beyond just weight fluctuation.

The HCG-801 is designed with precision and user convenience in mind. It features a large, easy-to-read LCD display that clearly presents the various measurements, making it accessible for users of all ages. The device can store data for multiple users, accommodating up to four individual profiles. This capability makes it suitable for families or shared fitness environments, allowing everyone to track their progress without mixing data.

To enhance the accuracy of measurements, the HCG-801 incorporates advanced sensor technology. This includes high sensitivity electrodes that ensure consistent readings regardless of factors such as hydration levels or time of day. Furthermore, the device has an ergonomic design, ensuring comfort during use and providing stability while users stand on the scale.

Connectivity features further elevate the user experience, as the HCG-801 can sync data with Omron's health management apps. This integration allows users to monitor their progress over time, set health goals, and receive personalized recommendations based on their body composition data. The Omron app is intuitive and provides users with graphs and trends to visualize their journey toward better health.

In terms of power efficiency, the HCG-801 runs on batteries, making it portable and easy to use at home or on the go. With its sleek design, durable construction, and reliable performance, the Omron HCG-801 stands out as an essential tool for individuals serious about their health and fitness. By providing detailed insights into body composition, it empowers users to make informed decisions regarding their exercise routines and dietary habits, promoting overall well-being.