Fluke 196, 199, 192 Triggering on External Waveforms, Select 1.2V under the Ext Level, Label

Page 58

Fluke 192/196/199

Users Manual

Triggering on External Waveforms

Use external triggering when you want to display waveforms on inputs A and B while triggering on a third signal. You can choose external triggering with automatic triggering or with edge triggering.

1Supply a signal to the red and black 4-mm banana jack inputs. See Figure 29.

In this example you continue from the Trigger on Edges example. To choose the external signal as trigger source, continue as follows:

2

 

 

 

 

Display the TRIGGER (On Edges)

 

 

 

 

 

 

 

 

 

key labels.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3 Select Ext (external) edge trigger.

Observe that the key labels at the bottom of the screen have been adapted to allow selection of two different external trigger levels: 0.12 V and 1.2 V:

 

 

 

Figure 29. External Triggering

 

 

 

 

 

 

 

4

 

 

 

 

 

Select 1.2V under the Ext LEVEL

 

 

 

 

 

 

 

 

 

 

 

label.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

From this point the trigger level is fixed and is compatible with logic signals.

50

Image 58
Contents Fluke 192/196/199 Page Limited Warranty & Limitation of Liability Service Centers Table of Contents Using Replay, Zoom and Cursors Using The MultimeterUsing The Recorder Functions Fluke 192/196/199Iii Specifications IndexStatement of Conformity Declaration of ConformityManufacturer Sample testsUnpacking the Test Tool Kit ScopeMeter Test Tool KitDescription # DescriptionSafety Information Read First Safety Information Read First Use of the test tool in a manner not specified may If Safety Features are ImpairedImpair the protection provided by the equipment Powering the Test Tool Using The ScopeAbout this Chapter Resetting the Test Tool Screen After ResetNavigating a Menu Open the Waveform OptionsHiding Key Labels and Menus Input ConnectionsMaking Scope Connections Scope ConnectionsDisplaying an Unknown Signal with Connect-and-View Screen After an Auto SetSelect the Peak-Peak Making Automatic Scope MeasurementsMeasurement Freezing the Screen Hz and V peak-peak as Scope ReadingsUsing Average for Smoothing Waveforms Using Average, Persistence and Glitch CaptureJump to Waveform Jump to Waveform and Select Using Persistence to Display WaveformsPersistence , then select Infinite Displaying Glitches Suppressing High Frequency NoiseSelecting AC-Coupling Acquiring WaveformsReversing the Polarity of the Displayed Waveform Select 10kHz HF reject to Mathematics menu Analyzing WaveformsMathematics... to open Or a vs BUsing The Multimeter Making Meter ConnectionsMaking Multimeter Measurements Measuring Resistance ValuesOpen the Current Probe Making a Current MeasurementHighlight a ac SubmenuAmpere Measurement Readings Freezing the Readings Selecting Auto/Manual RangesMaking Relative Measurements Making a Relative MeasurementUsing The Recorder Functions Opening the Recorder Main MenuStarting a TrendPlot Function Plotting Measurements Over Time TrendPlotHighlight Trend Plot Scope Displaying Recorded Data TrendPlot ReadingOpen the Recorder Options Changing the Recorder OptionsTurning Off the TrendPlot Display MenuStarting a Scope Record Function Recording Scope Waveforms In Deep Memory Scope RecordHighlight Scope Record Using ScopeRecord in Single Sweep Mode Jump to Display Glitches Select Yes, then jump to Mode Using Scope Record in Triggered Single Sweep ModeLevel Analyzing a TrendPlot or Scope Record Triggered Single Sweep RecordingUsing Replay, Zoom and Cursors Replaying the 100 Most Recent Scope ScreensReplaying Step-by-Step Replaying a WaveformTurning Off the Replay Function Replaying ContinuouslyCapturing 100 Intermittents Automatically Zooming in on a Waveform Zooming in a WaveformDisplaying the Zoomed Waveform Turning Off the Zoom FunctionUsing Horizontal Cursors on a Waveform Making Cursor MeasurementsUsing Vertical Cursors on a Waveform Turn off the cursorsUsing Cursors on a A+B, A-B or A*B Waveform Making Rise Time MeasurementsTriggering on Waveforms Setting Trigger Level and Slope Screen with all Trigger InformationUsing Trigger Delay or Pre-trigger Trigger Delay or Pre-trigger ViewOpen the Trigger Options menu Automatic Trigger OptionsOpen the Automatic Trigger Select Free Run, jump to Noise Triggering on Edges Open the Trigger Options menu Open the Trigger on Edge menu Reject FilterSelect On Trigger, jump to Noise Triggering on Noisy WaveformsMaking a Single Acquisition Set Noise reject Filter to OnMaking a Single Shot Measurement Select 1.2V under the Ext Level Triggering on External WaveformsLabel Triggering on Video Signals Triggering on Video Frames Enable video line selectionTriggering on Video Lines Detecting Narrow Pulses Triggering on PulsesSelect On Trigger Triggering on Narrow Glitches Finding Missing Pulses Open the Trigger Options menu Select On TriggerTriggering on Missing Pulses Using Memory, PC and Printer Saving and RecallingSaving Screens with Associated Setups Recalling Screens with Associated Setups Deleting Screens with Associated SetupsDisplay the SAVE/PRINT key labels Recalling a Setup Configuration Viewing Stored Screens Scroll through all stored screensConnecting to a Computer Documenting ScreensConnecting to a Printer Connecting a Serial PrinterPrinting a Screen Setting up the Printing ConfigurationOpen the Printer Setup submenu Baud RateFluke 192/196/199 Using the Standard Accessories TipsProbing Using 2-mm Heavy Duty Test Probes Heavy Duty Fixed Connections for ScopeTips Using the Standard Accessories Using the Independently Floating Isolated Inputs Measuring Using Independently Floating Isolated InputsTips Using the Independently Floating Isolated Inputs Turn the test tool off Using the Tilt StandSuppressing Key Labels and Menu’s Press and holdOpen the Language Select Adjusting the Contrast and BrightnessChanging the Information Language Accept French as the languageChoose 1999, jump to Month Open Date Adjust menuChanging Date and Time Choose 04, jump to DaySaving Battery Life Setting the Power Down TimerTo Coupling Open the User Options menu Open the Auto Set Adjust MenuChanging the Auto Set Options Select UnchangedMaintaining the Test Tool Cleaning the Test ToolStoring the Test Tool Charging the Batteries Charging the BatteriesExtending Battery Operation Time Open the User Options menuSelect 101, then return Replacing the NiMH Battery Pack BP190Calibrating the Voltage Probes Select Voltage, then jump toMaintaining the Test Tool Displaying Calibration Information Parts and AccessoriesStandard Accessories Open the Version & CalibrationStandard Accessories Ordering CodeTL24 General Purpose Leads RS200 Optional Accessories Troubleshooting Battery Operated Fluke Accessories Do Not Function FlukeView Does Not Recognize The Test ToolSafety Characteristics SpecificationsPerformance Characteristics IntroductionIsolated Inputs a and B Vertical Dual Input OscilloscopeHorizontal Edge Trigger Trigger and DelayAutomatic Connect-and-View Trigger Isolated External TriggerAutomatic Scope Measurements AC Voltage VAC AC+DC Voltage True RMSFrequency Hz Amperes AMPPeak Duty Cycle DutyTemperature Temp PowerPhase Decibel dBMeter DMM Measurements on Meter InputsWith Optional Temperature Probe DiodeTrendPlot Meter or Scope RecorderScope Record Zoom, Replay and Cursors MiscellaneousOptical InterfacePort Probe CalibrationMemory MechanicalEnvironmental Max. Input Voltages SafetyMax. Floating Voltage Probe SafetyElectrical specifications Environmental102 Electromagnetic Immunity Disturbance less than 10% of full scale = 3V/m104 Index Numerics106 107 108