DECOMPRESSION DIVE, ABOVE CEILING. NOTE DOWNWARD POINTING ARROW, ER WARNING AND ALARM. YOU SHOULD IMMEDIATLY (WITHIN
3 MINUTES) DESCEND TO OR BELOW CEILING.
In addition, an Error warning (Er) reminds you that you have only three minutes to correct the situation. You must immediately descend to, or below, the ceiling.
If you continue to violate the decompression, the dive computer will go into a permanent Error Mode. In this mode, the instrument can only be used as a depth gauge and timer. You must not dive again for at least 48 hours. (See Section 5.7. Error conditions).
6.2. Diving in MIXED GAS mode
MIXED GAS mode is the first dive mode available in Suunto HelO2. The mode is used when diving with air or with oxygen- or
6.2.1. Before diving in MIXED GAS mode
If set to the MIXED GAS mode, the correct oxygen and helium percentage of the gas in your cylinders must always be entered into the computer to ensure correct inert gas and oxygen calculations. The dive computer adjusts its mathematical inert gas and oxygen calculation models accordingly. The dive computer will not accept fractional percentage values of oxygen and helium concentration. Do not round up fractional percentages. For example, 31.8% oxygen should be entered as 31%. Rounding up will cause inert gas percentages to be understated and will affect decompression calculations. If you want to adjust the computer to provide more conservative calculations, use the personal adjustment feature to affect decompression calculations, or reduce the PO2 setting to affect oxygen exposure according to the entered O2% and PO2 values. Calculations based on nitrox use result in longer
As a safety precaution, the oxygen calculations in the computer are made with an oxygen percentage of 1% + set O2% .
When the dive computer is set in MIXED GAS mode, the Dive Planning mode calculates using the O2% and PO2 values that are currently in the computer.
To set the nitrox, trimix and/or or heliox mixes, refer to Section 5.9.1. Setting the gases.
Default gas mix settings
In MIXED GAS mode, the Suunto HelO2 allows you to set 1 – 8 gas mixes containing 8 – 99% oxygen and 0 – 92% helium.
In the MIXED GAS mode, the default setting is standard air (21% O2 and 0% He). It remains in this setting until the O2% is adjusted to any other percentage of oxygen (8% – 99%). The default setting for maximum oxygen partial pressure is 1.4 bar, however you are able to set it in the range of 0.5 - 1.6 bar.
6.2.2. Oxygen and helium displays
When MIXED GAS mode is activated, the display will show the information in the figure below. In MIXED GAS mode, the maximum operational depth is calculated based on set O2%, He% and PO2 values.
31