Agilent Technologies 60501B Voltage 3 to 60 V minimum dc operation from 0 to 2 V for 0 to 30 A

Page 4
Voltage: 3 to 60 V (minimum dc operation from 0 to 2 V for 0 to 30 A)

Table 60501-1. Specification and Supplemental Characteristics

SPECIFICATIONS

DC Input Rating:

Current: 0 to 30 A

Voltage: 3 to 60 V (minimum dc operation from 0 to 2 V for 0 to 30 A)

Power: 150 W at 40 °C (derated to 112 W at 55 °C)

A. OPERATING CHARACTERISTICS

B. DERATED CURRENT DETAIL

Constant Current Mode:

 

 

Ranges:

0 to 3 A; and 0 to 30 A

Accuracy:

(after 30 second wait): ± 0.1% ± 40 mA (both ranges)

Resolution:

0.8 mA (3 A range); 8 mA (30 A range)

Regulation:

10 mA (both ranges)

Temperature Coefficient:

100 ppm/°C ± 3 mA/°C (both ranges)

Constant Resistance Mode:

 

 

Ranges:

0.067 to 2 Ω; 2 Ω to 2 kΩ; and 20 Ω to 10 kΩ

Accuracy:

± 0.8% ± 16 m Ω with 6 A at input (2 Ω range);

 

± 0.3% ± 5 mS with 6 V at input (2 k and 10 kΩ ranges)

Resolution:

0.54 mΩ (2 Ω range); 0.14 mS (2 kΩ range); 0.014 mS (10 kΩ range)

Regulation:

10 mV with remote sensing (2 Ω range); 10 mA (2 k and 10 kΩ ranges)

Temperature Coefficient:

800 ppm/°C ± 0.8 mΩ/°C (2 Ω range);

 

300ppm/°C ± 0.5 mS/°C (2 k and 10 kΩ ranges)

Constant Voltage Mode:

 

 

Range:

0 to 60 V

 

Accuracy:

± 0.1% ± 50 mV

Resolution:

16 mV

 

Regulation:

5 mV (remote sense); 40 mV (local sense)

Temperature Coefficient:

100 ppm/°C ± 5 mV/°C

2

Image 4
Contents OPERATING MANUAL 150 WATT ELECTRONIC LOAD MODULE Agilent Model 60501BPrinted in U.S.A DECLARATION OF CONFORMITY according to ISO/IEC Guide 22 and ENSafety Manufacturer’s NameModule Installation and Operation About This ManualItems Supplied 150-Watt ModuleVoltage 3 to 60 V minimum dc operation from 0 to 2 V for 0 to 30 A 6% of setting ± 2% 0.25 Hz to 10 kHz3% to 97% 0.25 Hz to 1 kHz 6% to 94% 1 kHz to 10 kHz 50 μs ± 3% minimum 4 s ± 3% maximum10 kHz 3 db frequency Range CRNG value CURRRANG valueTable 60501-3. Factory Default Settings SettingsSetting Table 60501-2. Programming Ranges continuedPower Supply Settings Table 60501-4. Calibration InformationRanges and Calibration Points VariablesAgilent Sales and Support Office United StatesLatin America Canada