Agilent Technologies 60501B manual CRNG value, CURRRANG value

Page 8
C:RNG value

Table 60501-1 Specifications and Supplemental Characteristics (continued)

Source Turn-On Current Overshoot: Less than 10% of final value (in CC and CR modes when connected to power supplies with voltage rise times of greater than 500μs).

Programmable Short Circuit: 0.066 Ω (0.04 Ω typical)

Programmable Open Circuit: 20 kΩ (typical)

Drift Stability (over an 8 hour interval):

 

 

Current:

± 0.03%

± 5 mA

Voltage:

± 0.01%

± 10 mV

Reverse Current Capacity:

50 A when unit is on; 20 A when unit is off

Weight:

Function

Constant Current

Set Range

Low Range

High Range

3.2 kg (7 lbs.)

Table 60501-2. Programming Ranges

Front Panel

Front Panel

HPSL Command

Key

Display

(Short Form)

 

C:RNG value

"CURR:RANG value"

 

 

 

 

Range of Values

0 and 3 A

> 3 A and 30 A

Set Main Level

Low Range

High Range

Set Slew Rate

Low Range

High Range

Set Transient Level

*Set Triggered Level

Constant Resistance

Set Range

Low Range

Middle Range

High Range

Set Main Level

Low Range

Middle Range

High Range

Set Slew Rate

Low Range

Middle/High Range

Set Transient Level

*Set Triggered Level

(shift) "CURR:RANG value"

(shift) Manual background

CURR value

C:SLW value

C:TLV value

R:RNG value

RES value

V:SLW value C:SLW value R:TLV value

"CURR value"

"CURR:SLEW value"

"CURR:TLEV value" "CURR:TRIG value"

"RES:RANG value"

"RES value"

"VOLT:SLEW value" "CURR:SLEW value"

"RES:TLEV value" "RES:TRIG value"

0 to 3 A

0 to 30 A

0.00005 to 0.25 (A/μs)

0.0005 to 2.5 (A/μs) same as main level same as main level

0 and 2 Ω

>2 Ω and 2 kΩ >2 kΩ and 10 kΩ

0 to 2 Ω

2 Ω to 2 kΩ

20 Ω to 10 kΩ

same as voltage slew same as current slew same as main level same as main level

Constant Voltage

Set Main Level

Set Slew Rate

Set Transient Level

*Set Triggered Level

VOLT value

(shift)

V:SLW value

V:TLV value

"VOLT value"

"VOLT:SLEW value" "VOLT:TLEV value" "VOLT:TRIG value"

0 to 60 V

0.001 to 0.5 (V/μs) same as main level same as main level

6

Image 8
Contents Printed in U.S.A 150 WATT ELECTRONIC LOAD MODULE Agilent Model 60501BOPERATING MANUAL DECLARATION OF CONFORMITY according to ISO/IEC Guide 22 and ENSafety Manufacturer’s NameModule Installation and Operation About This ManualItems Supplied 150-Watt ModuleVoltage 3 to 60 V minimum dc operation from 0 to 2 V for 0 to 30 A 6% of setting ± 2% 0.25 Hz to 10 kHz3% to 97% 0.25 Hz to 1 kHz 6% to 94% 1 kHz to 10 kHz 50 μs ± 3% minimum 4 s ± 3% maximum10 kHz 3 db frequency Range CRNG value CURRRANG valueTable 60501-3. Factory Default Settings SettingsSetting Table 60501-2. Programming Ranges continuedPower Supply Settings Table 60501-4. Calibration InformationRanges and Calibration Points VariablesAgilent Sales and Support Office United StatesLatin America Canada
Related manuals