7.Attach the other end of the air tubing to the base of the nebulizer.

NOTE: Ensure nebulizer tubing is firmly attached to compressor before use.

WARNING

In the case of high humidity, condensation may form inside the tubing. Run the unit for two minutes before connecting the nebulizer to the tubing.

8.Attach the mouthpiece to the top section of the nebulizer. Refer to the nebulizer manual for additional instructions.

NOTE: Before starting a cold unit, let the unit reach room temperature in a heated area.

9.With the power switch off, plug the power cord into an appropriate electrical outlet.

CAUTION

This device is for intermittent use: 20 min. ON / 40 min. OFF. Failure to operate within these limits could damage the device beyond repair.

Invacare® Pediatric Bear

18

Part No 1160837

Image 18
Contents Invacare Pediatric Bear Nebulizer System Invacare Pediatric Bear Table of Contents Register Your Product Special Notes WordImportant Safeguards To Reduce the Risk of Electrocution Important Safeguards Important Safeguards Important Safeguards Introduction Typical Product Parameters Typical Product Parameters Supplemental Information CompressorExtension Cords Unpacking HandlingShipping Instructions InspectionSetup Operating InstructionsAdding Medication Operating Instructions Using Treatment Instructions for NebulizerTreatment Instructions for Mask Cleaning and Disinfecting Cleaning the Outer CaseCleaning the Nebulizer and Mask Cleaning and Disinfecting Cleaning the Compressor Cleaning the MouthpieceMaintenance Routine CareStorage Nebulizer Change Filter Change Troubleshooting Troubleshooting Limited Warranty Limited Warranty Limited Warranty USA