SECTION 4—MAINTENANCE

Nebulizer Change

The nebulizer is designed for single patient use. The nebulizer MUST be replaced after a long period of inactivity, or when it is obstructed by dry medication, dust, etc. Reusable nebulizers should last from six months to one year with general use. Refer to the nebulizers owner’s manual. Disposable nebulizers should last 30 days. For disposable nebulizers, it is recommended that an extra nebulizer be kept at all times. Refer to the disposable nebulizer’s owner’s manual.

Filter Change

1.The filter MUST be replaced after approximately 30 hours of use or when it turns grey.

2.Open the filter holder to replace the filter with a new one.

CAUTION

DO NOT use cotton or any other material.

DO NOT wash or clean the filter.

Use only Invacare filters.

DO NOT operate unit without an air filter.

Invacare® Pediatric Bear

26

Part No 1160837

Image 26
Contents Invacare Pediatric Bear Nebulizer System Invacare Pediatric Bear Table of Contents Register Your Product Special Notes WordImportant Safeguards To Reduce the Risk of Electrocution Important Safeguards Important Safeguards Important Safeguards Introduction Typical Product Parameters Typical Product Parameters Extension Cords Supplemental InformationCompressor Unpacking HandlingShipping Instructions InspectionSetup Operating InstructionsAdding Medication Operating Instructions Using Treatment Instructions for NebulizerTreatment Instructions for Mask Cleaning and Disinfecting Cleaning the Outer CaseCleaning the Nebulizer and Mask Cleaning and Disinfecting Cleaning the Compressor Cleaning the MouthpieceStorage MaintenanceRoutine Care Nebulizer Change Filter Change Troubleshooting Troubleshooting Limited Warranty Limited Warranty Limited Warranty USA