SECTION 3—CLEANING AND DISINFECTING

DANGER

DO NOT attempt to separate the outer case of this unit. Removing the case creates a serious electrical shock hazard. All disassembly and maintenance of this unit MUST be done by a qualified technician.

Cleaning the Nebulizer and Mask

WARNING

To prevent possible risk of infection due to contamination, follow these instructions:

Clean the nebulizer and mask before initial operations.

Clean the nebulizer and mask after each aerosol treatment.

ALWAYS use fresh cleaning solution.

DO NOT towel dry the nebulizer components or mask after cleaning.

DO NOT boil the mask, damage will occur.

NOTE: Your physician and/or an Invacare dealer may specify certain cleaning procedures. If so, follow their recommendations. If not, clean according to the following instructions:

Invacare® Pediatric Bear

22

Part No 1160837

Image 22
Contents Invacare Pediatric Bear Nebulizer System Invacare Pediatric Bear Table of Contents Register Your Product Special Notes WordImportant Safeguards To Reduce the Risk of Electrocution Important Safeguards Important Safeguards Important Safeguards Introduction Typical Product Parameters Typical Product Parameters Compressor Supplemental InformationExtension Cords Unpacking HandlingShipping Instructions InspectionSetup Operating InstructionsAdding Medication Operating Instructions Using Treatment Instructions for NebulizerTreatment Instructions for Mask Cleaning and Disinfecting Cleaning the Outer CaseCleaning the Nebulizer and Mask Cleaning and Disinfecting Cleaning the Compressor Cleaning the MouthpieceRoutine Care MaintenanceStorage Nebulizer Change Filter Change Troubleshooting Troubleshooting Limited Warranty Limited Warranty Limited Warranty USA