5. Cleaning and daily disinfecting

5.1Cleaning

Following the cleaning instructions after each use will prevent any remaining medication in the tank from drying, resulting in the device not nebulising effectively and also for preventing infections.

Warning:

Thoroughly rinse the nebuliser kit and the mask, nosepiece or mouthpiece with clean hot tap water after each use, and disinfect the parts of the nebuliser kit and applicable inhalation attachments after the last treatment of the day.

Note: Do not start operation before the cleaned parts are completely dried.

1.Make sure that the power switch is in the off ( ) position.

2.Disconnect the AC adapter from the wall outlet and the compressor.

3.Remove the nebuliser kit from the air tube and disassemble the nebuliser kit. Refer to Section 3 (page 11-13).

4.Discard any remaining medication in the

medication tank.

5.Wash all the nebulising parts in warm water and mild detergent and rinse in clean hot tap water.

For instructions on how to disinfect the parts, see the following section.

6.After cleaning and disinfecting the parts, rinse them thoroughly with clean water and allow the parts to air dry in a clean environment.

EN

17

Image 17
Contents Model NE-C801 Contents Intended use IntroductionUsage Important safety instructionsImportant safety instructions Save these instructions for future reference General Safety PrecautionsNebuliser Kit Inhalation Attachments Know your unitNebuliser Kit Compressor Main Unit Accessories Preparing the Nebuliser for Use Guide to using the devicePreparing the nebuliser for use Connecting AC adapterCapacity of the medication tank is 2 to 7 ml Adding medication and assembling the partsAttaching the mouthpiece Preparing the nebuliser for use Into the mouth Taking medication Direction. Medication may flow Using the nosepiece Taking medication Cleaning and daily disinfecting Compressor is not waterproof Disinfecting the nebulising partsAlways wash the vaporiser head after each Handling the Vaporiser HeadParts Model Materials An example of commercially available disinfectant Maintenance and storage Changing the air filter Problem Cause Remedy TroubleshootingContents Technical dataTechnical data Technical data Technical data Child Mask Sebs Set Air Filters x AC Adapter Optional parts and accessoriesMade in China