8. Troubleshooting
Check the following if your unit should fail during operation. You can also refer to the pages of this manual for complete instructions.
Problem | Cause | Remedy | |
|
|
| |
| Is the AC adapter connected | Check that the plug is | |
Nothing happens when the | inserted in an electrical | ||
power switch is pressed. | correctly to the an electrical | outlet. Unplug then reinsert | |
outlet and the compressor? | |||
| the plug if necessary. | ||
|
| ||
|
|
| |
| Is there medication in the |
| |
| medication tank? | Add the correct amount of | |
| Is there too much/little | medication to the medication | |
| medication in the medication | tank. | |
| tank? |
| |
| Is the vaporiser head missing | Attach the vaporiser head | |
| and not assembled correctly? | correctly. | |
| Is the nebuliser kit | Assemble the nebuliser kit | |
| assembled correctly? | correctly. | |
|
|
| |
| Is the nozzle blocked? | Make sure that the nozzle is | |
No nebulisation or low | free of blockages. | ||
| |||
nebulisation rate, when the |
|
| |
Is the nebuliser kit tilted at a | Make sure that the nebuliser | ||
power is on. | kit is not tilted at an angle of | ||
| sharp angle? | more than 45 degrees. | |
|
| ||
|
|
| |
| Is the air tube connected | Make sure that the air tube is | |
| correctly connected to the | ||
| correctly? | ||
| compressor and nebuliser kit. | ||
|
| ||
|
|
| |
| Is the air tube folded or | Make sure that the air tube | |
| damaged? | does not contain kinks. | |
|
|
| |
| Is the air tube blocked? | Make sure that the air tube is | |
| free of blockages. | ||
|
| ||
|
|
| |
| Is the air filter dirty? | Replace the air filter with a | |
| new one. | ||
|
| ||
The device is abnormally | Is the air filter cover attached | Attach the air filter cover | |
loud. | correctly? | correctly. | |
|
| Do not cover the compressor | |
| Is the compressor covered? | with any type of cover during | |
|
| use. | |
The device is very hot. |
|
| |
Is the device operating | Limit use to 20 minutes at a | ||
| time, and allow a 40 minute | ||
| continuously over 20 | ||
| interval before using the | ||
| minutes? | ||
| device again. | ||
|
| ||
|
|
|
Note: If the suggested remedy does not solve the problem, do not try to repair the device - no parts of the unit are user serviceable.
Return the unit to an authorized OMRON retail outlet or distributor.