Introduction

Thank you for purchasing the OMRON COMP AIR.

This product was developed in conjunction with respiratory therapists for the successful treatment of asthma, chronic bronchitis, allergies and other respiratory disorders. The compressor forces air to the nebuliser kit. When the air enters the nebuliser kit, it converts the prescribed medication into an aerosol of microscopic droplets that can easily be inhaled.

This is a medical device. Operate the device only as instructed by your doctor and/or respiratory therapist.

Intended use

Medical Purpose

This product is intended to be used for inhaling medication

 

for respiratory disorders.

 

Intended User

• Legally certified medical experts, such as doctor, nurse and

 

therapist, or healthcare personnel or patient under the

 

guidance of qualified medical experts.

 

• The user should also be capable of understanding general

 

operation of NE-C801 and the content of instruction

 

manual.

 

Intended Patients

This product should not be used by patients, who are

 

unconscious or are not breathing spontaneously.

Environment

This product is intended for use in a medical facility, such as

 

Hospital, clinic and doctor’s office, and in a room of general

 

household.

 

Durable period

Durable periods are as follows, provided the product is used

 

to nebulise 2 ml of medication 2 times a day for 6 minutes

 

each time at room temperature (23°C).

 

Durable period may vary depending on usage environment.

 

Compressor (Main unit)

5 years

 

Nebuliser Kit

1 year

 

(Cap, Inhalation Top,

 

 

Vaporiser Head, Medication Tank)

 

 

Mouthpiece

1 year

 

Nosepiece

1 year

 

Air Tube (PVC, 100cm)

1 year

 

Air Filter

60 days

 

Adult Mask (PVC)

1 year

 

Child Mask (PVC)

1 year

Precautions for use Warnings and cautions described in the instruction manual should be observed.

EN

3

Image 3
Contents Model NE-C801 Contents Intended use IntroductionUsage Important safety instructionsImportant safety instructions Save these instructions for future reference General Safety PrecautionsKnow your unit Nebuliser Kit Compressor Main UnitNebuliser Kit Inhalation Attachments Accessories Preparing the Nebuliser for Use Guide to using the devicePreparing the nebuliser for use Connecting AC adapterCapacity of the medication tank is 2 to 7 ml Adding medication and assembling the partsAttaching the mouthpiece Preparing the nebuliser for use Taking medication Direction. Medication may flowInto the mouth Using the nosepiece Taking medication Cleaning and daily disinfecting Compressor is not waterproof Disinfecting the nebulising partsAlways wash the vaporiser head after each Handling the Vaporiser HeadParts Model Materials An example of commercially available disinfectant Maintenance and storage Changing the air filter Problem Cause Remedy TroubleshootingContents Technical dataTechnical data Technical data Technical data Child Mask Sebs Set Air Filters x AC Adapter Optional parts and accessoriesMade in China