Omron ne-c801 instruction manual Taking medication

Page 16

4.Taking medication

4.When treatment is complete, turn the power off. Check that there is no condensation or moisture in the air tube and unplug the device from the power outlet.

Warning:

Condensation may occur in the air tube. Do not store the air tube while it contains condensation or moisture. This could result in infection via bacteria.

If there is moisture or liquid remaining in the air tube, be sure to follow the procedure below to remove any moisture from within the air tube.

1)Make sure that the air tube is still connected to the air connector on the compressor.

2)Unplug the air tube from the nebuliser kit.

3)Turn on the compressor and pump air through the air tube to expel the moisture.

16

Image 16
Contents Model NE-C801 Contents Introduction Intended useImportant safety instructions UsageImportant safety instructions General Safety Precautions Save these instructions for future referenceNebuliser Kit Compressor Main Unit Know your unitNebuliser Kit Inhalation Attachments Accessories Guide to using the device Preparing the Nebuliser for UseConnecting AC adapter Preparing the nebuliser for useAdding medication and assembling the parts Capacity of the medication tank is 2 to 7 mlAttaching the mouthpiece Preparing the nebuliser for use Direction. Medication may flow Taking medicationInto the mouth Using the nosepiece Taking medication Cleaning and daily disinfecting Disinfecting the nebulising parts Compressor is not waterproofHandling the Vaporiser Head Always wash the vaporiser head after eachParts Model Materials An example of commercially available disinfectant Maintenance and storage Changing the air filter Troubleshooting Problem Cause RemedyTechnical data ContentsTechnical data Technical data Technical data Optional parts and accessories Child Mask Sebs Set Air Filters x AC AdapterMade in China
Related manuals
Manual 32 pages 36.76 Kb Manual 30 pages 40.59 Kb