4.Cleaning and daily disinfecting

4.2Disinfecting the nebulising parts

Always clean before disinfection. (See Section 4.1 “Cleaning”)

Always disinfect the medication tank, mouthpiece and other nebulising parts after the last treatment of the day. For appropriate method of disinfecting each part, refer to page 18.

To disinfect the parts, use one of the methods described below:

A. Use a commercially available disinfectant. Follow the instructions provided by the disinfectant manufacturer.

1.Submerge the parts in the cleansing solution for the specified period.

2.Remove the parts and discard the solution.

3.Rinse the parts with clean hot tap water, shake off excess water and allow to air dry in a clean environment.

B.Parts may be boiled between 15 to 20 minutes.

After boiling, carefully remove the parts, shake off excess water and allow to air dry in a clean environment.

Note: Do not boil air tube, mask, kids accessory, air filter and air filter cover.

Caution:

Do not use an autoclave, EOG gas disinfection or low temperature plasma sterilizer to disinfect the device.

Note: The printing may come off the surfaces of kids accessories as a result of repeated washing and disinfecting.

Handling the Vaporiser Head

Baffle

 

The vaporiser head is an important part that is used to vaporise medication.

Be sure to observe the following precautions when handling it.

Caution:

• Always wash the vaporiser head after each use.

Do not use a brush or pin etc., to clean the parts.

When disinfecting the parts by boiling, be sure to boil them in plenty of water.

• Do not boil vaporiser head together with other objects, other than

Vaporiser Head

applicable nebuliser accessories.

 

• Do not use a microwave oven, dish dryer or hair dryer to dry parts.

EN

17

Image 17
Contents Model NE-C801KD Quick Manual Connecting AC adapter Introduction ContentsIntended use Usage Important safety instructionsUsage Save these instructions for future reference General Safety PrecautionsNebuliser Kit Compressor Main Unit Know your unitAccessories Preparing the nebuliser for use Connecting AC adapterAdding medication and assembling the parts Page Hold the nebuliser kit as indicated on the right Taking medicationPage Cleaning and daily disinfecting Disinfecting the nebulising parts Always clean before disinfection. See .1 CleaningAn example of commercially available disinfectant Maintenance and storage Changing the air filter Problem Cause Remedy TroubleshootingName Model Type Rating AC adapter Technical dataTechnical data Technical data Air Tube PVC, 100cm Air Tube Silicon, 100cm Optional parts and accessoriesOmron Santé France SAS