7. Troubleshooting

Check the following if your unit should fail during operation. You can also refer to the pages of this manual for complete instructions.

Problem

Cause

Remedy

 

 

 

Nothing happens when the

Is the AC adapter connected

Check that the plug is

power switch is pressed.

correctly to the an electrical

inserted in an electrical

 

outlet and the compressor?

outlet. Unplug then reinsert

 

 

the plug if necessary.

No nebulisation or low

Is there medication in the

Add the correct amount of

nebulisation rate, when the

medication tank?

medication to the medication

power is on.

 

tank.

Is there too much/little

 

 

 

medication in the medication

 

 

tank?

 

 

Is the vaporiser head missing

Attach the vaporiser head

 

and not assembled correctly?

correctly.

 

Is the nebuliser kit

Assemble the nebuliser kit

 

assembled correctly?

correctly.

 

Is the nozzle blocked?

Make sure that the nozzle is

 

 

free of blockages.

 

Is the nebuliser kit tilted at a

Make sure that the nebuliser

 

sharp angle?

kit is not tilted at an angle of

 

 

more than 45 degrees.

 

Is the air tube connected

Make sure that the air tube is

 

correctly?

correctly connected to the

 

 

compressor and nebuliser kit.

 

Is the air tube folded or

Make sure that the air tube

 

damaged?

does not contain kinks.

 

Is the air tube blocked?

Make sure that the air tube is

 

 

free of blockages.

 

Is the air filter dirty?

Replace the air filter with a

 

 

new one.

The device is abnormally

Is the air filter cover attached

Attach the air filter cover

loud.

correctly?

correctly.

The device is very hot.

Is the compressor covered?

Do not cover the compressor

 

 

with any type of cover during

 

 

use.

 

Is the device operating

Limit use to 20 minutes at a

 

continuously over 20

time, and allow a 40 minute

 

minutes?

interval before using the

 

 

device again.

Note: If the suggested remedy does not solve the problem, do not try to repair the device - no parts of the unit are user serviceable.

Return the unit to an authorized OMRON retail outlet or distributor.

EN

21

Image 21
Contents Model NE-C801KD Quick Manual Connecting AC adapter Introduction ContentsIntended use Usage Important safety instructionsUsage Save these instructions for future reference General Safety PrecautionsNebuliser Kit Compressor Main Unit Know your unitAccessories Preparing the nebuliser for use Connecting AC adapterAdding medication and assembling the parts Page Hold the nebuliser kit as indicated on the right Taking medicationPage Cleaning and daily disinfecting Disinfecting the nebulising parts Always clean before disinfection. See .1 CleaningAn example of commercially available disinfectant Maintenance and storage Changing the air filter Problem Cause Remedy TroubleshootingName Model Type Rating AC adapter Technical dataTechnical data Technical data Air Tube PVC, 100cm Air Tube Silicon, 100cm Optional parts and accessoriesOmron Santé France SAS