Omron NE- C801KD instruction manual Adding medication and assembling the parts

Page 12

2.Preparing the nebuliser for use

2.2Adding medication and assembling the parts

1. Remove the mask or mouthpiece, kids accessory and cap from the nebuliser kit.

2. Remove the inhalation top from the medication tank.

1) Rotate the inhalation top anti-clockwise.

2) Lift the inhalation top out of the medication tank.

3. Add the correct amount of prescribed medication to the medication tank.

Caution:

The capacity of the medication tank is 2 to 7 ml.

Note: Scales on the nebuliser kit are for reference only. The scale on the outside of the tank is with no vaporiser head in place. Please use the scale on your syringe or vial for accurate measurement of medication.

4.Make sure that the vaporiser head is securely

attached to the inhalation top.

5. Put the inhalation top back inside the medication tank.

1) Align the protrusion on the inhalation top with the indent on the medication tank, as shown.

2) Rotate the inhalation top clockwise and align with the medication tank until they click in place.

12

Image 12
Contents Model NE-C801KD Quick Manual Connecting AC adapter Contents IntroductionIntended use Important safety instructions UsageUsage General Safety Precautions Save these instructions for future referenceKnow your unit Nebuliser Kit Compressor Main UnitAccessories Connecting AC adapter Preparing the nebuliser for useAdding medication and assembling the parts Page Taking medication Hold the nebuliser kit as indicated on the rightPage Cleaning and daily disinfecting Always clean before disinfection. See .1 Cleaning Disinfecting the nebulising partsAn example of commercially available disinfectant Maintenance and storage Changing the air filter Troubleshooting Problem Cause RemedyTechnical data Name Model Type Rating AC adapterTechnical data Technical data Optional parts and accessories Air Tube PVC, 100cm Air Tube Silicon, 100cmOmron Santé France SAS