Possible problems and how to solve them
Symptom |
| Check | Remedy | |||
Operation button |
| ➀ Check that the plug of the AC | Correctly insert it. | |||
lamp (AC) does |
| adaptor is inserted correctly. |
|
|
| |
not light up. |
| ➁ Check that the main unit is | Correctly mount it. | |||
• With AC |
| mounted correctly onto the AC |
|
|
| |
adaptor in use. |
| adaptor. |
|
|
| |
|
| ➂ Check that the level of water in | Fill water up to proper level. | |||
|
| the water tank is normal. |
|
|
| |
|
|
|
|
| ||
No nebulisation |
| ➀ Check that standard volume of | Fill the medication cup and switch | |||
|
| the medication is in the cup. | on the power by pressing OFF/ON | |||
|
|
|
| ( |
| ). |
|
|
|
|
| ||
|
|
|
|
| ||
|
|
|
|
| ||
Low nebulisation |
|
|
| Replace the medication cup. | ||
| ➀ Be sure the medication cup is | |||||
rate |
| not broken. |
|
|
|
|
|
| ➁ Check the medication level in |
| Remove the exess medication. | ||
|
| cup. |
|
|
|
|
|
| ➂ Check aerosol volume adjust |
| Increase delivery rate. | ||
|
| knob. |
|
|
|
|
|
| ➃ Check battery light. |
| Go to AC operation. | ||
|
| ➄ Check preparation of |
| Eventually read information sup- | ||
|
| medication. |
| plied with medication or consult | ||
|
|
|
| your doctor. | ||
|
|
|
|
|
|
|
If you find unsolvable problems please contact your authorised OMRON Customer Service. For address see packaging and/or attached literature.
English
Technical data
This device fulfills the provisions of the EC directive 93/42/EEC (Medical Device Directive).
Model: | OMRON NE U07 | ||||||||
Electrical rating: | AC 220V – 230V~, 50/60 Hz | ||||||||
Power consumption: | Approx. 20 Watts | ||||||||
Particle size: | 3.5 µm (MMD *) | ||||||||
Ultrasonic frequency: | 2.4 MHz | ||||||||
Nebulisation rate: | Approx. 1.0 ml/min. | ||||||||
Water tank: | Approx. 21 ml | ||||||||
Fuse: |
|
|
| = T 0.1 A | |||||
|
|
| |||||||
Medication cup: | maximum 10 ml | ||||||||
Dimensions: | Approx. 62 mm (W) x 110 mm (D) x 254 mm (H) | ||||||||
Weight: | approx. 940 g | ||||||||
Operating conditions: | 0 °C – 40 °C; 30% RH – 85% RH | ||||||||
Storage conditions: | 0 °C – 40 °C; 30% RH – 85% RH | ||||||||
Electrical safety: |
|
|
|
|
| - Class II, | |||
|
|
|
|
|
| - Type B | |||
Symbol explanation: |
|
|
|
|
|
|
|
| = Manual operation |
|
|
|
|
|
| ||||
|
|
|
|
|
| ||||
| AC = Alternating current | ||||||||
|
|
|
|
|
| = OFF/ON | |||
|
|
|
| ||||||
|
|
|
|
Technical data subject to change
Manufacturer:
OMRON HEALTHCARE Europe
* Mass Median Diameter
Data independently obtained by: Malvern Instruments GmbH, Germany on Malvern Master Sizer System
14 | 15 |