WHAT DISPLAY SYMBOLS MEAN
Display | Condition/ | Recommended |
Symbol | Cause | Action |
❤Appears during measure- ment; blinks and beeps
MEASURING when pulse is detected.
Remain still and refrain from talking; measurement in progress.
ERROR
Appears on display prior to | Press exhaust button to |
and during measurement. | release air. |
Blinks if there is air in cuff |
|
at power up. |
|
|
|
| Measurement begun, but | Increase pressurization by | |
|
|
| pressure is not sufficient for | 30 mmHg to 40 mmHg and | |
reading. | repeat measurement. | ||||
|
|
|
|
| |
|
|
| Battery voltage is | Replace all batteries | |
|
|
| |||
|
|
| too low for monitor | with new ones. | |
|
|
| |||
|
|
| |||
LOW BATTERY | to work properly. |
| |||
|
| ||||
|
|
|
|
| |
|
|
| Appears when blood pres- | Exhaust air from cuff, reap- | |
|
|
| sure cannot be measured | ply cuff properly, power up | |
| ERROR | accurately due to erratic | again and repeat measure- | ||
| pulse or noise interference. | ment in quiet environment. | |||
|
|
|
|
| |
|
|
| Appears when pulse cannot | Exhaust air from cuff, | |
|
|
| be measured accurately. | reapply cuff properly, power | |
|
|
|
| up again and repeat | |
PULSE ERROR |
| measurement. | |||
|
|
|
|
| |
| EXHAUST | Occurs when exhaust | Adjust exhaust rate to 2 | ||
| velocity rate is more than 8 | mmHg to 5 mmHg. | |||
SPEED FLASHING | |||||
mmHg. |
| ||||
|
|
|
| ||
|
|
|
|
| |
|
|
| No Display Symbol | Check for correct air con- | |
|
|
|
| nector. Make sure air con- | |
|
|
|
| nector is inserted correctly | |
|
|
|
| into the hoses and monitor. |
CUFF HOSE | BULB HOSE |
MONITOR