4.Troubleshooting and Maintenance

4.1The Icons and Error Messages

Error Display

Cause

Remedy

 

 

 

 

 

Remove the arm cuff. Wait 2-3 minutes and then

 

Irregular or weak pulses are detected.

take another measurement.

 

Repeat the steps in section 3.3. If this error

 

 

 

 

continues to appear, contact your doctor.

 

 

 

 

Movement during measurement.

Carefully read and repeat the steps in section 3.3.

 

 

 

 

Cuff is not applied correctly.

Apply the arm cuff correctly. Refer to section 3.1.

 

 

 

 

 

You should replace them with new ones ahead of

Blink

The batteries are low.

time.

 

 

Refer to section 2.1.

Lit

The batteries are exhausted.

You should replace them with new ones at once.

Refer to section 2.1.

 

 

 

 

 

12

Image 14
Contents Model M3 Contents Important Safety Information General UsageImportant Safety Information Overview Main unitInstalling/Replacing Batteries PreparationSetting the Buzzer/ Date and Time Press the O/I Start button to store the settingUsing the Unit Applying the Arm CuffHow to Sit Correctly Cuff wrapping guideTaking a Reading Press the O/I Start buttonUsing the Unit Using the Memory Function To delete all the values stored in memory Troubleshooting and Maintenance Icons and Error MessagesTroubleshooting and Maintenance Troubleshooting Problem Cause RemedyTroubleshooting and Maintenance Maintenance Calibration and ServiceStorage Place the arm cuff and main unit in the storage caseOptional Parts Medium Arm Cuff Large Arm Cuff AC Adapter STechnical Data Power SourceTechnical Data Technical Data Some Useful Information about Blood Pressure Why is it a Good Thing to measure Blood Pressure at Home? Some Useful Information about Blood Pressure