6.Technical Data

Product Description

Digital Automatic Blood Pressure Monitor

Model

OMRON M3 (HEM-7200-E(V)/ HEM-7200-E8(V))

Display

LCD Digital Display

Measurement Method

Oscillometric method

Measurement Range

Pressure: 0 mmHg to 299 mmHg

 

Pulse: 40 to 180/min.

Accuracy

Pressure: ±3 mmHg

 

Pulse: ± 5% of display reading

Inflation

Fuzzy-logic controlled by electric pump

Deflation

Automatic pressure release valve

Memory

60 Measurements with date and time

Rating

DC6V

4W

Power Source

4 “AA” batteries 1.5V or AC/ DC adapter

 

(Adapter S-9515336-9 (optional), INPUT AC100-240V 50/60Hz 0.12A)

Battery Life

Capacity of new alkaline batteries is approx. 1500 measurements

Applied Part

 

 

 

= Typeype BB

Protection Against Electric

Internally powered ME equipment (When using only the batteries)

Shock

 

 

 

 

 

 

 

 

= Class II ME equipment (Optional AC adapter)

Operating temperature/ Humidity +10°C to +40°C / Maximum: 30 to 90% RH

Storage temperature/ Humidity/

-20°C to +60°C / Maximum: 10 to 95% RH / 700-1060hPa

Air pressure

 

Console Weight

Approximately 340g without batteries

Cuff Weight

Approximately 120g

Outer Dimensions

Approximately 123 (w) mm × 85 (h) mm × 141(l) mm

Cuff Dimensions

Approximately 146 mm × 446 mm (Medium cuff: arm circumference 22 to 32 cm)

Cuff Material

Nylon and polyester

Package Content

Main unit, cuff, instruction manual, storage case, battery set, guarantee card, blood

 

pressure pass

Note: Subject to technical modification without prior notice.

EN

19

Image 21
Contents Model M3 Contents General Usage Important Safety InformationImportant Safety Information Main unit OverviewPreparation Installing/Replacing BatteriesPress the O/I Start button to store the setting Setting the Buzzer/ Date and TimeApplying the Arm Cuff Using the UnitCuff wrapping guide How to Sit CorrectlyPress the O/I Start button Taking a ReadingUsing the Unit Using the Memory Function To delete all the values stored in memory Icons and Error Messages Troubleshooting and MaintenanceTroubleshooting and Maintenance Problem Cause Remedy TroubleshootingTroubleshooting and Maintenance Calibration and Service MaintenancePlace the arm cuff and main unit in the storage case StorageMedium Arm Cuff Large Arm Cuff AC Adapter S Optional PartsPower Source Technical DataTechnical Data Technical Data Some Useful Information about Blood Pressure Why is it a Good Thing to measure Blood Pressure at Home? Some Useful Information about Blood Pressure