Omron M24/7 Not required to repeat with each session, Preparation of Monitoring Session

Page 13

A successful monitoring session consists of the following steps:

1.Connecting the OMRON M24/7 monitor to the PC (First connection only

– not required to repeat with each session)

1.Connect the optical interface, which was delivered with the device, to the USB port of the PC.

2.The other side of the optical USB interface cable, which is a two-point plastic connection, should be connected to the socket of the OMRON M24/7, in a way that the red plastic ring is directed to the lower side of the device.

3.Start the OMRON BP Tracker Software and open the Options menu, then click on the Communication tab on the left side of the screen.

4.On the appearing window, click on the USB option. Please connect the device to the PC (fresh/charged batteries, device switched on) as noted above then click on the Test button. Upon successful communication the software will display the serial number and the rmware version of the device.

2.Preparation of Monitoring Session

1.Inform your patient about the monitoring rules well in advance.

2.Programming the measurement plan into the device:

Using PC: start the OMRON BP Tracker Software program

Optionally the device can be programmed without using a PC, for details see Manual Programming section.

3.Enter new patient data or select patient from the database.

4.Create a monitoring plan with respect to the patient’s lifestyle.

5.Insert two fresh or fully charged, AA size batteries into the battery compartment and check their voltage.

6.Connect the monitor to the computer.

7.Send the monitoring plan from the computer to the monitor unit.

8.Apply the cuff to the patient with the device placed in the carry pouch.

9.Give the Patient’s Quick Guide to the patient along with detailed instructions about the rules and the usage of the device.

3.Evaluation

1.Remove the unit and cuff from the patient on his/her return.

2.Ask for the Patient’s Quick Guide /patient diary, and ask the patient for any events, symptoms, observations or complaints.

3.Start the OMRON BP Tracker Software

4.Connect the device to the PC and then transfer collected data from the monitor to your database.

5.Analyze the blood pressure pro le.

6.Create and print a report.

13

Image 13
Contents M24/7 Ambulatory Blood Pressure Monitor Introduction Content Important Safety Information Important Safety InformationDamage to the equipment or other property Indications Recommended use of ambulatory blood pressure monitoringContraindications Main Unit OverviewArm Cuffs Front sideDisplay Package ContentsCancel a blood pressure measurement Using the ButtonsOperating Instructions Working with the Omron M24/7 UnitSwitching the device on Switching the device offBattery voltage check LCD checkSet by PC Overview DisplayMonitoring Step by Step Rules of MonitoringPreparation of Monitoring Session Not required to repeat with each sessionEvaluation Start the Omron BP Tracker SoftwareProgramming Options Manual ProgrammingMeasurement Timing Manual Programming Step by StepSoftware BatteriesCuffs and their application Handling Errors and Problems Care and MaintenanceProtection, cleaning and washing Accessories Maintenance and Storage ConditionsRegular checks, warranty, service Correct Disposal of this ProductTechnical parameters Technical DataElectric shock hazard protection Safety ConcernsBiocompatibility Hazardous materialsEmission test Compliance Electromagnetic Environment EMC InformationPage Recommended Separation Separation distance in function of the frequency Recommended separation distance=3.5/E Transmitter W Ambulatory Blood Pressure Monitoring Product Warranty Information Remarks