Omron M24/7 Recommended separation distance, Separation distance in function of the frequency

Page 25

Recommended separation distance

OMRON M24/7 is intended to be used in electromagnetic environment with controlled RF disturbances. The purchaser or user of M24/7may help to reduce electromagnetic disturbances by de ning the separation distance between the transportable or mobile RF telecommunication equipment (transmitters) and OMRON M24/7, depending on the highest output power of the telecommunication equipment.

 

Separation distance in function of the frequency of the

 

transmitter [m]

 

 

 

The highest

150KHz –

80MHz –

 

800MHz – 2.5GHz

output

80MHz

800MHz

 

d=[7/E1] P

power of the

d=[3.5/V1]

d=[3.5/E

] P

 

transmitter [W]

 

1

 

 

 

 

 

 

0.01

Not applicable

0.12

 

0.23

0.1

Not applicable

0.38

 

0.73

1

Not applicable

1.2

 

2.3

10

Not applicable

3.8

 

7.3

100

Not applicable

12

 

23

If this table does not contain the highest output power of the transmitter, the d separation distance [m] can be calculated by the formula, depending on the frequency of the transmitter, where P is the rated highest output power of the transmitter [W].

1.note: in case of frequency 80MHz or 800 MHz, the formula for the higher range is applicable.

2.note: These are guidelines. Actual conditions may vary.

25

Image 25
Contents M24/7 Ambulatory Blood Pressure Monitor Introduction Content Important Safety Information Important Safety InformationDamage to the equipment or other property Indications Recommended use of ambulatory blood pressure monitoringContraindications Main Unit OverviewArm Cuffs Front sideDisplay Package ContentsCancel a blood pressure measurement Using the ButtonsOperating Instructions Working with the Omron M24/7 UnitSwitching the device on Switching the device offBattery voltage check LCD checkSet by PC Overview DisplayMonitoring Step by Step Rules of MonitoringPreparation of Monitoring Session Not required to repeat with each sessionEvaluation Start the Omron BP Tracker SoftwareProgramming Options Manual ProgrammingMeasurement Timing Manual Programming Step by StepSoftware BatteriesCuffs and their application Handling Errors and Problems Care and MaintenanceProtection, cleaning and washing Accessories Maintenance and Storage ConditionsRegular checks, warranty, service Correct Disposal of this ProductTechnical parameters Technical DataElectric shock hazard protection Safety ConcernsBiocompatibility Hazardous materialsEmission test Compliance Electromagnetic Environment EMC InformationPage Recommended Separation Separation distance in function of the frequency Recommended separation distance=3.5/E Transmitter W Ambulatory Blood Pressure Monitoring Product Warranty Information Remarks