Recommended separation distance
OMRON M24/7 is intended to be used in electromagnetic environment with controlled RF disturbances. The purchaser or user of M24/7may help to reduce electromagnetic disturbances by de ning the separation distance between the transportable or mobile RF telecommunication equipment (transmitters) and OMRON M24/7, depending on the highest output power of the telecommunication equipment.
| Separation distance in function of the frequency of the | ||||
| transmitter [m] |
|
|
| |
The highest | 150KHz – | 80MHz – |
| 800MHz – 2.5GHz | |
output | 80MHz | 800MHz |
| d=[7/E1] P | |
power of the | d=[3.5/V1] | d=[3.5/E | ] P | ||
| |||||
transmitter [W] |
| 1 |
|
| |
|
|
|
| ||
0.01 | Not applicable | 0.12 |
| 0.23 | |
0.1 | Not applicable | 0.38 |
| 0.73 | |
1 | Not applicable | 1.2 |
| 2.3 | |
10 | Not applicable | 3.8 |
| 7.3 | |
100 | Not applicable | 12 |
| 23 |
If this table does not contain the highest output power of the transmitter, the d separation distance [m] can be calculated by the formula, depending on the frequency of the transmitter, where P is the rated highest output power of the transmitter [W].
1.note: in case of frequency 80MHz or 800 MHz, the formula for the higher range is applicable.
2.note: These are guidelines. Actual conditions may vary.