4. Handling Errors and Problems
4.2Troubleshooting
| Problem | Cause | Remedy | |
|
| Is your arm placed | Refer to Chapter 3.2. | |
|
| through the arm cuff | ||
|
| correctly? |
| |
|
|
| Keep quiet during | |
|
| Talking during the | measurement. | |
|
| Carefully read and repeat | ||
|
| measurement. | ||
|
| the steps listed under | ||
The reading is |
| |||
| Chapter 3.3. | |||
extremely high (or low). |
| Carefully read and repeat | ||
|
| Are rolled up sleeves | ||
|
| the steps listed under | ||
|
| squeezing the arm? | ||
|
| Chapter 3.3. | ||
|
|
| ||
|
| Is the cuff lining in the | Cancel the measurement | |
|
| and contact your | ||
|
| arm cuff damaged? | OMRON distributor or | |
|
|
| Customer Services. | |
|
| Is air leaking from the | Contact your OMRON | |
Cuff is not inflated. | distributor or Customer | |||
arm cuff? | ||||
|
| Services. | ||
|
|
| ||
The unit operates normally and measurements |
| |||
appear to be taken, but... |
|
| ||
• | Measurements taken by a doctor are higher (or | Refer to Chapter 8. | ||
| lower). |
| EN | |
• | Measurement results are different each time. |
| ||
|
| The AC adapter was | Connect the AC adapter. | |
|
| disconnected from the | ||
|
| (Refer to Chapter 2.1.) | ||
No display appears | unit. | |||
| ||||
when O/I button is | The AC adapter was | Connect the AC adapter | ||
pressed. | ||||
|
| disconnected from the | to the power supply. | |
|
| power supply. | (Refer to Chapter 2.1.) | |
Readings are not stored | G (guest) was selected, | Select the correct user | ||
or no user ID was | ID. | |||
in the memory. | ||||
selected. | (Refer to Chapter 3.3.) | |||
|
|
37