OS1592 Infrared Fiber Optic Thermometer/Transmitter 2
2.2 – Electrical Connection
Attach the optical assembly to the optical adapter located on the side of the housing. Refer to Fig. 1 for the overall appearance of the unit.
Open the cover of the aluminum housing. Slide your cable through the metal feed thru and connect the wires to the 9 position terminal block (J101) as shown in Fig. 2. Depending on the type of the analog outputs, you need to make the proper connection. Connect the shield of the cable to the inside of the metal feed thru to minimize RF noise.
In order to use the Backlight source
ALARM LED
INDICATION
TO THE ASSEMBLY
® | FIBER OPTIC |
| THERMOMETER |
|
|
HAL | MODE SET |
POWER
OPTICAL
CONNECT
CURRENT TEMPERATURE
°F
ALARM
OS1590 SERIES
AND OUTPUT
CONNECTIONS
4 POSITION MEMBRANE KEYPAD
MODEL & S/N
Figure 1 - OS1592 Main Unit
9
RELAY/NO
RELAY/NC |
| RELAY |
| ||
| CONTACTS |
| |||
|
|
|
| ||
RELAY/COM |
| ||||
– |
|
|
|
|
|
|
|
|
|
| |
+ | mA 4/20mA |
| |||
+ | 1mV/Deg | mV |
|
| |
+ |
|
|
| ||
| 0/5VDC, 0/10VDC | V/TC | |||
|
| J & K T/C | |||
|
|
|
–
+
1
J101
ANALOG OUTPUTS
POWER SUPPLY
Figure 2 - Power Supply and Analog Output Connections
3