ReliOn Thermometer Taking a Reading, Changing Modes, Recalling Measurements in Memory

Page 5

Taking a Reading

NOTE: Many pediatricians recommend the use of ear thermometers for children older than 6 months.

1. Press and release the SCAN button. After two seconds,

 

the system will perform a self-test, Fig. 1, and show the

Fig. 1

last temperature taken along with a flashing “°F” to indicate

the unit is ready for measurement, Fig. 2.

2. Gently insert the probe snugly into the ear canal until a

slight resistance is felt. The thermometer will beep when Fig. 2 the auto-position sensor detects proper placement. If the

unit does NOT beep, gently reposition the thermometer.

NOTE: To disable the auto-position sensor, immediately press the SCAN button after inserting the thermometer into the ear canal.

3. After the unit beeps, press the SCAN button, Fig. 3, and

 

Fig. 3

 

 

promptly release the button. The device will beep once

 

 

 

 

again after the measurement is complete.

 

 

NOTE: If you do NOT press the SCAN button once it is

 

 

 

 

 

 

 

placed in the ear canal, the thermometer will automatically

 

Fig. 4

take a measurement in 9 seconds.

 

 

4. If the measurement is below 100.4°F (38°C), the

 

 

thermometer will beep once and the measurement will be

 

 

 

 

 

 

 

shown on the display with a happy face symbol, Fig. 4.

 

Fig. 5

5.If the temperature is 100.4°F (38°C) or above, the thermometer will sound one long beep followed by two short beeps. The measurement will be shown on the display with a sad face symbol to indicate fever, Fig. 5.

6.Thermometer will automatically beep and shut off after approximately 1 minute of non-use.

NOTE: To ensure accuracy,

1.Please wait at least 10 seconds between successive readings.

2.Clean the probe lens before and after each use.

Changing Modes

Temperature readings can be displayed in either Fahrenheit or Celsius. To switch between °F or °C, press and hold the SCAN button. Without releasing, also press the ‘M’ button

and toggle between the modes, Fig. 1 and 2.

 

 

 

Fig. 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fig. 2

 

 

 

 

 

Recalling Measurements in Memory

 

 

 

 

 

 

 

 

 

 

The last 10 measurements taken are automatically

 

 

 

 

 

 

 

 

 

 

 

 

stored into memory. To recall measurements, press

 

 

 

 

 

 

the SCAN button to turn the unit on. After the unit

 

 

 

 

 

 

beeps, press the ‘M’ button to view measurements in

Memory

Index

Fig. 1

memory. Continue to press the ‘M’ button to view the

 

 

 

Symbol

 

 

 

 

 

 

 

next previously stored measurement, Fig. 1. The unit will beep each time a recorded measurement is viewed.

Replacing the Battery

Your Instant Ear Thermometer requires one lithium battery, type CR2032. The battery needs to be replaced when the low battery symbol appears on the display.

To replace the battery:

1) Slide the battery cover off.

2) Using a non-metal pointed tool, carefully remove the battery and insert the new battery with the positive (+) side facing up.

3) Replace the battery cover.

NOTE: Please properly dispose the battery away from small children and heat.

8

9

Image 5
Contents Instant Ear Thermometer Index Limited Five-Year WarrantyHelpful Hints Mabis HealthcareNite-GloTMDisplay Product FeaturesIntroduction To Clean Cleaning and StorageGeneral Care & Important Information To StoreTaking a Reading Replacing the BatteryChanging Modes Recalling Measurements in MemoryDisplay Symbol Condition/Cause Corrective Action SymbolsTroubleshooting Product Specifications Sugerencias Útiles Garantía Limitada de Cinco AñosÍndice Exhibición de Nite-GloTM IntroducciónFuciones de Producto Limpieza y Almacenamiento Cuidado General e Información ImportantePara Limpiar Para GuardarCómo Cambiar la Batería Cómo Tomar una LecturaCómo Cambiar de Modo Símbolo Condición/Causa Acción Correctiva Solución de ProblemsSímbolos Especificaciones del Producto