DESCRIPTION OF DISPLAY MARKS
DISPLAY MARK | CONDITION/CAUSE | CORRECTIVE ACTION | ||||||||
|
|
|
|
| ||||||
SYSTOLIC mmHg |
| DIASTOLIC mmHg | Measurement is begun, but the pressure | Change the pressurization setting one level by pressing | ||||||
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
| was insufficient. | “Pressure Setting Switch” and measure once again. |
AVE. | 1 | 2 | 3 | 4 | 5 | 6 | 7 | PULSE/min |
| (Automatic |
|
|
| ||||||||
|
|
|
|
| ||||||
SYSTOLIC mmHg |
| DIASTOLIC mmHg | Appears when the blood pressure could |
| ||||||
|
|
|
|
|
|
|
|
| not be measured accurately. (Some |
|
|
|
|
|
|
|
|
|
| noise was detected or moved arm dur- | Repeat measurement again. |
AVE. | 1 | 2 | 3 | 4 | 5 | 6 | 7 | PULSE/min | ing measuring.) | |
|
|
| ||||||||
|
|
|
| First exhaust air from the cuff, reapply the cuff properly | ||||||
SYSTOLIC mmHg |
| DIASTOLIC mmHg |
| |||||||
|
|
|
|
|
|
|
|
| Appears when the pulse could not ac- | and remeasure. |
|
|
|
|
|
|
|
|
| curately be measured. |
|
AVE. | 1 | 2 | 3 | 4 | 5 | 6 | 7 | PULSE/min |
|
|
|
|
| ||||||||
|
|
|
|
| ||||||
SYSTOLIC mmHg |
| DIASTOLIC mmHg |
|
| ||||||
|
|
|
|
|
|
|
|
| Appears when the batteries become | Replace the batteries. |
|
|
|
|
|
|
|
|
| weak. | |
|
|
|
|
|
|
|
| PULSE/min |
| |
AVE. | 1 | 2 | 3 | 4 | 5 | 6 | 7 |
|
| |
|
|
| ||||||||
|
|
|
|
| ||||||
SYSTOLIC mmHg |
| DIASTOLIC mmHg |
| Take a measurement. | ||||||
|
|
|
|
|
|
|
|
| Appears when no memory is stored at | |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
| PULSE/min | pressing of “Memory Recall Switch” | Blood pressure will store automatically. |
AVE. | 1 | 2 | 3 | 4 | 5 | 6 | 7 |
| ||
|
|
| ||||||||
|
|
|
|
| ||||||
SYSTOLIC mmHg |
| DIASTOLIC mmHg |
|
| ||||||
|
|
|
|
|
|
|
|
| Units has a trouble. | Ask purchased shop. |
AVE. | 1 | 2 | 3 | 4 | 5 | 6 | 7 | PULSE/min |
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
- Eng 12 -