Graham Field JB0112-164 user manual Troubleshooting, Problem Cause and Solution

Page 11

Storage

Keep the device and accessories dry. Avoid exposure to direct sunlight.

TROUBLESHOOTING

If your Neb-u-Tyke RR Train Nebulizer Compressor fails to function, consult the following Troubleshooting guide. If the problem persists, consult your Graham-Field autho- rized distributor.

Problem

Cause and Solution

 

 

Device doesn’t

Check if plug is properly fit into a grounded electrical outlet.

operate

 

When device has been run continuously for more than 30 minutes

 

 

just before using, auto-shut-down may have been activated by the

 

built-in thermal protector; ensure that device is OFF for at least 30

 

minutes before next usage.

 

 

Nebulization is

Check for proper electrical voltage.

weak

 

Check tubing for blockage or air leakage at connection to Neb-u-Tyke

 

 

compressor or nebulizer cup; replace as needed.

 

 

 

Check that the nebulizer cup is properly assembled and not

 

damaged. If there is any damage, replace as needed.

 

 

 

Check if filter is dirty; replace filter as needed.

 

 

John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 11

Image 11
Contents Neb-u-TykeRR Train Contents Introduction Symbols KeyImportant Safety Precautions John Bunn Neb-u-Tyke RR Train Nebulizer Compressor Product Description SET-UP Operation Maintenance Cleaning the Nebulizer Cup Problem Cause and Solution TroubleshootingSpecification Description SpecificationsLimited Warranty John Bunn Neb-u-Tyke RR Train Nebulizer Compressor GF Health Products, Inc. Graham-Field GF Health Products, Inc Manual de Usuario Contenidos Símbolos IntroducciónDirectivas Importantes DE Seguridad John Bunn Neb-u-Tyke RR Train Compresor de Nebulizador Descripción DEL Producto Ensamblaje Operación Mantenimiento Limpieza de la Taza del Nebulizador Problema Causa y Solución Detectando ProblemasEspecificación Descripción EspecificacionesGarantía John Bunn Neb-u-Tyke RR Train Compresor de Nebulizador GF Health Products, Inc. Graham-Field JB0112-164-INS-LAB-RevB09