Invacare WT 200 Attaching the Walking Tutor to the Lift, Removing the Boom Bearing Retainer

Page 3

Socket Head Screws

Allen Key

DETAIL “A”

 

DETAIL “B”

 

 

 

 

Actuator and Boom

Bearing Block

 

Mast

 

 

Mast

 

 

 

 

Kneepad

 

 

 

DETAIL “C”

Boom Bearing

Retainer

 

 

Mast

FIGURE 2 Removing the Boom Bearing Retainer

Attaching the Walking Tutor to the Lift

NOTE: For this procedure, refer to FIGURE 3 on page 4.

1.Install the pivot assembly to the mast.

2.Replace the boom bearing retainer.

3.Swing the boom back and push the bearing block into the top of the mast.

4.Secure the bearing block to the top of the mast with the existing socket head screws.

5.Ensure that the walking tutor armrests are positioned the correct way before assembling (not shown).

6.Position the walking tutor in behind the actuator.

7.Rotate the walking tutor to match the boom orientation.

8.Using the socket head screws and locknuts, secure the walking tutor bearing block to the pivot assembly (Detail “C” and Detail “D”).

Part No 1141439

3

Walking Tutor

Image 3
Contents Attaching Strap Assemblies to the Boom Safety SummaryAssembling the Walking Tutor Walking TutorAttaching Strap Assemblies to the Boom Removing the Boom Bearing RetainerRemoving the Boom Bearing Retainer Attaching the Walking Tutor to the LiftAttaching the Walking Tutor to the Lift Assembling the Armrest to Walking TutorDepth Adjusting the ArmrestHeight AngleUsing the Walking Tutor Removing the FootplateFitting the Sling to a Patient Lifting the Patient Positioning the Walking Tutor to Suit the PatientWalking the Patient Product Made Australia