Tanita BD-815U instruction manual Output Data Format, Explanation for RS-232C-compliant output

Page 12

Output Data Format

Explanation for RS-232C-compliant output

This explanation is for the interface to output the readings from the GB Digital Baby-Scale BD-815U to external equipment (e.g. PC) as a RS-

232C-compliant signal.

Grams (g) mode (transmission data)

Transmission data is output immediately after measurement regard- less of the status of the reception side (e.g. PC). Therefore, the recep- GB tion side must always be ready to receive data before any measure- ments are taken.

Specifications

Communication standard

EIA RS-232C

Communication method

Asynchronous

 

 

Signal speed

9,600bps

Data bit length

8bit

 

 

Parity

NONE

Stop bit

1bit

 

 

Terminator

CR+LF

Flow control

None

 

 

Signal line name and connection method

Terminal no.Signal name

1

 

2

TXD

3

 

4

 

5

GND

6

 

7

 

8

 

9

 

Connection example

BD-815U

 

 

PC, etc.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A. When the display is held

 

 

Item

 

 

Header

 

Output data (ASCII code)

Details

 

 

 

 

 

 

 

 

 

Control data

 

 

{0

 

Fixed at 16

 

 

Control data

 

 

~0

 

Fixed at 1

 

 

 

 

 

 

 

 

 

Model No.

 

MO

 

"XXXXXX"

8-byte fixed length (“BD-815”)

 

Body weight

 

Wg

 

XXXXX

3~5 bytes variable length (unit: g)

 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

5

6

7

8

9

10

11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27

{ 0 , 1 6 , ˜ 0 , 1 , M O , " B D - 8 1 5 " , W g , 3

28 29 30

31

32

9 7 2 CR LF

B. When the Weight Gain key is pressed

 

 

Item

 

 

Header

 

Output data (ASCII code)

Details

 

 

 

 

 

 

 

 

 

 

 

 

 

Weight gain

 

 

Gg

 

XXXXXX

 

 

 

 

1~6 bytes variable length (unit: g)

Memorized weight

 

Rg

 

XXXXX

 

 

 

 

3~5 bytes variable length (unit: g)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

5

6

7

8

9

10

11 12 13 14

15

 

16

 

 

G g 5 , 1 2 0 , R g , 2 8 6

CR

 

LF

 

 

C. When the Recall key is pressed

 

 

Item

 

 

Header

Output data (ASCII code)

Details

 

 

 

 

 

 

 

 

 

Memorized weight

 

Rg

XXXXX

3~5 bytes variable length (unit: g)

1

2

3

4

5

6

7

8

 

 

9

 

 

R

g

,

3

8

6

8

CR

 

 

LF

 

 

Note

•Commas (,) are used to separate data.

•Terminator (end of data) is CR (ASCII code: 0DH), LF (ASCII code: 0AH).

22

23

Image 12
Contents Professional Baby Scale BD-815UContents Maintenance Safety NotesSafety Notes Instructions for MeasurementBaby Basket Attachment & Removal AttachmentRemoval Confirmation of accessories FrontDisplay Panel BackUsing batteries Level gauge CheckWhen tare is not specified Measurement instructionsExtended Display Function Canceling the tare function Tare weighing functionDisplay Weight Gain Memory cancellationMemory Function Memory Value Recall MethodVarious Setting Methods Auto Power Off TimeDisplay Hold Time When the display is held When the Weight Gain key is pressedWhen the Recall key is pressed Output Data FormatOutput Data Format GB Problem Check pointsTanita Corporation. All Rights Reserved BD8157001