Guidance and manufacturer’s declaration—electromagnetic immunity

The C-Series Tango system is intended for use in the electromagnetic environment specified below. The customer or the user of the C-Series Tango system should assure that the device is used in such an environment.

Immunity

IEC60601-1-2

Complianc

Electromagnetic environment—

test

test level

e level

guidance

 

 

 

 

 

 

 

Portable and mobile RF communications

 

 

 

equipment should be used no closer to any

 

 

 

part of the C-Series Tango system, including

 

 

 

cables, than the recommended separation

 

 

 

distance calculated from the equation

 

 

 

applicable to the frequency of the transmitter.

 

 

 

Recommended separation distance

Conducted RF

3 Vrms

3 Vrms

d = 1.17 P

IEC 61000-4-6

150 kHz to 80 MHz

 

 

Radiated RF

10 V/m

10 V/m

d = 0.35 P 80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2.5 GHz

 

 

 

 

 

d = 0.70 P 800 MHz to 2.5 GHz

 

 

 

where P is the maximum output power rating

 

 

 

of the transm tter in watts (W) according to the

 

 

 

transmitter manufacturer and d is the

 

 

 

recommended separation distance in meters

 

 

 

(m).

 

 

 

Field strengths from fixed RF transm tters, as

 

 

 

determined by an electromagnetic site

 

 

 

survey,a should be less than the compliance

 

 

 

level in each frequency range.b

 

 

 

Interference may occur in the vicin ty of

 

 

 

equipment marked with the following symbol:

 

 

 

 

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all s tuations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transm tters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the C-Series Tango system is used exceeds the applicable RF compliance level above, the C-Series Tango system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the C-Series Tango system.

bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

English

13

Image 15
Contents Series Tango Manufacturer Distributed by Intended Use ContraindicationsAdverse Effects Masks Parts of the TangoTango Interface Setting up the TangoUsing the Tango Weekly Cleaning and MaintenanceChanging the Air Filter TroubleshootingPower Supply Technical SpecificationsSymbols which appear on the product General Warnings and Cautions Servicing English Page Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty ROW ENG
Related manuals
Manual 2 pages 20.9 Kb

C-Series specifications

ResMed C-Series is a line of advanced medical devices designed to provide effective therapy for patients suffering from obstructive sleep apnea (OSA) and other respiratory conditions. Known for its cutting-edge technology and patient-friendly features, the C-Series stands out in the realm of continuous positive airway pressure (CPAP) machines.

One of the main features of the ResMed C-Series is its AutoSet technology, which automatically adjusts the pressure levels throughout the night based on the patient’s needs. This responsiveness ensures optimal therapy comfort and minimizes disturbances that can disrupt sleep. Patients benefit from a personalized treatment experience that adapts as their conditions change.

The C-Series also incorporates a quiet operation mode, significantly reducing noise levels compared to traditional CPAP machines. This is particularly beneficial for couples, allowing for peaceful sleep without the distraction of machine noise. Additionally, the C-Series boasts an intuitive user interface with a bright LCD screen, making it easy for patients to navigate settings and access key information about their therapy.

Comfort is a priority in the design of the C-Series. The device supports a range of mask options, including nasal and full-face masks, allowing patients to choose the most comfortable fit for their individual needs. Moreover, the built-in HumidAir integrated humidifier helps maintain optimal humidity levels during therapy, reducing discomfort associated with dry air.

Another cornerstone of the C-Series is its advanced data tracking capabilities. The device records essential metrics such as adherence, mask fit, and leak levels, which can be accessed easily through the ResMed myAir app. This connectivity empowers patients to take an active role in their care, while also enabling healthcare providers to monitor progress remotely.

Durability and portability are additional characteristics of the C-Series. Its compact design makes it travel-friendly, allowing users to maintain their therapy routine while on the go. The device is also built with high-quality materials to ensure long-lasting performance, making it a wise investment for long-term health.

In summary, the ResMed C-Series is a leading choice for those seeking effective sleep apnea treatment. With its innovative features, user-centric design, and advanced technology, patients can expect a comfortable, quiet, and responsive therapy experience that can significantly improve their quality of life.