Omega TX82B manual Configuration Procedures, Calibration, FACTORY-SETUP Label

Page 14

6.2CONFIGURATION PROCEDURES

1.Remove all push-on jumpers.

2.For an input current range of 1-5 mA, no jumper is required. For 4-20 mA input, install a push-on jumper at S1-H.

For 10-50 mA input, install a push-on jumper at S1-J.

3.If N2 (Section 6.1) is less than N1, reverse the signal polarity by removing jumpers from S1-B and S1-C and reinstalling jumpers at S1-A and S1-D.

4.Select the zero range required (ZR1-4) and install the push-on jumper as indicated in the configuration chart (Section 5.2).

5.If a decimal point is required (DP1-5), install a push-on jumper as indicated in the configuration chart (Section 5.2).

6.If one dummy right-hand zero is required, install jumper at S2-A.

7.If two dummy right-hand zeros are required, install jumpers at S2-A, S2-G, S2-H, and S2-J.

6.3CALIBRATION

1.Apply an input current (I1) and adjust the zero pot (Z) to read N1.

2.Apply an input current (I2) and adjust the span pot (S) to read N2.

3.Repeat steps 1 and 2 as required to set N1 and N2 to within ±1 count.

6.4FACTORY-SETUP LABEL

The label on the meter shows the factory-configured input and display settings. If the configuration is changed, use the extra label included with the meter to indicate the new settings.

10

Image 14
Contents TX82B Servicing North America USATable of Contents Unpacking Instructions QTY DescriptionGeneral Information Accuracy AT 25C SpecificationsInput ConversionMechanical DisplayEnvironment Mechanical Assembly and Installation Safety ConsiderationsINSTALLATION/REMOVAL Exploded View includes rear panelSignal Input Connections P3 Field Configuration Chart ConfigurationStandard Setup Configuration Jumper InstallationSpan ADJ Zero ADJ VR2 Customer Configuration and Calibration FormulaCalibration Configuration ProceduresFACTORY-SETUP Label Wiring Diagram Wiring Diagram for TX82B Usage for FM Factory MutualPage Page Page WARRANTY/DISCLAIMER Temperature