Omega TX82B manual Specifications, Input, Accuracy AT 25C, Conversion

Page 6

2.0SPECIFICATIONS

2.1INPUT

 

Current

1-5 mA, 4-20 mA or 10-50 mA

 

Protection

200 mA max forward and 1000 mA max

 

 

reverse

 

Voltage Drop

2.5 V max forward and 1.2 V max reverse

 

Span Range

100 to 2000 counts continuous adjustment

 

 

with a multiturn potentiometer

 

Zero Range

The multiturn zero potentiometer can displace

 

 

the displayed reading by ±500 counts from the

 

 

center of each of the four overlapping zero

 

 

ranges, provided that the resultant is within the

 

 

±1999 count display capability.

 

Span Slope

Positive or Negative

 

Normal Mode Rejection

 

 

at 50/60 Hz

46 dB minimum

2.2

ACCURACY AT 25°C

 

 

Maximum error

±0.1% R ±1 count

 

Zero tempco

±0.1 ct/K typ, ±0.3 ct/K max

 

Span tempco

±0.005%S/K typ, ±0.015%S/K max

2.3

CONVERSION

 

 

Technique

Dual-slope, average value with autozero

 

 

correction

 

Polarity

Determined automatically at the end of input

 

 

integration period

 

Input integration period

100 milliseconds (nominal value)

 

Reading rate

2.5/second

2

Image 6
Contents TX82B Servicing North America USATable of Contents Unpacking Instructions QTY DescriptionGeneral Information Accuracy AT 25C SpecificationsInput ConversionDisplay EnvironmentMechanical Mechanical Assembly and Installation Safety ConsiderationsINSTALLATION/REMOVAL Exploded View includes rear panelSignal Input Connections P3 Field Configuration Chart ConfigurationStandard Setup Configuration Jumper InstallationSpan ADJ Zero ADJ VR2 Customer Configuration and Calibration FormulaConfiguration Procedures FACTORY-SETUP LabelCalibration Wiring Diagram Wiring Diagram for TX82B Usage for FM Factory MutualPage Page Page WARRANTY/DISCLAIMER Temperature