6.0CUSTOMER CONFIGURATION AND CALIBRATION
Use this procedure to determine the configuration of the TX82B customized setup. The procedure is general; customers can specify any two current inputs and their corresponding digital readings.
6.1FORMULA
Base all your calculations on either the
1. Determine the lowest input current, I1, which is specified by the customer:
I1 = |
| mA |
2.Determine the highest input current, I2, which is specified by the customer:
I2 = |
| mA |
3.Determine the reading, N1, at input current I1, which is specified by the customer.
N1 = |
| counts |
4.Determine the reading, N2, at input current I2, which is specified by the customer:
| N2 = |
| counts | ||||
5. Calculate the Gain, G1: |
|
| |||||
| (N2 - N1) |
|
| ||||
G1 = |
| = |
|
| counts per mA | ||
(I2 - I1) | |||||||
|
|
(If G1 is greater than 125, it is out of range for a standard unit.)
6.Calculate the Required Zero Range number, RZR: RZR = N1 - (I1 x G1) =
7.Select the Zero Range required from the following chart where the Required Zero Range number falls between the upper and lower limits of that range:
ZR1 =
ZR2 =
ZR3 =
ZR4 = +760 to +2000
ZR =